Error
Description and solution
Error 12
:
Error 16
The CO
2
concentration sensor is not available or does not
work correctly.
Send the Multilogger for repair.
Error 19
The source value for the calculated channel is not available.
Check calculated channel settings and its source values.
Error 20
Calculation error, incorrectly calibrated device.
Send the Multilogger for repair.
Error 21
Counter overflow.
Erase the counter using the SW and repeat measurement.
If you do not want to be considered the counter overflow as
error you can change this setting of the input.
Error 22
Internal voltage error. Undesirable coupling among the
inputs, incorrect connection, damaged cables, damaged
device.
Check the input connections and the connected cables.
Error 50,
Error 53
:
Error 56
Device configuration error, damaged configuration.
Download the configuration to the SW, check of the settings
of individual items and save them back to the Multilogger.
Error 51
The probe line contains invalid calibration constants.
Send the probe for repair or contact the manufacturer.
Error 52
The probe of the Pt1000/M line contains invalid calibration
constants.
Send the probe for repair.
Error 90
Error 91
The Ethernet interface - the measured value is out of
numerical range.
Verify the measured value on the device display, contact
the manufacturer.
Error 128
:
Error 255
Communication error.
Repeat the communication, contact the manufacturer.
Содержание AR004926
Страница 5: ...5 Multilogger drawing only certain models...
Страница 10: ...10 The way of mounting the holder to Multilogger and mounting holes dimensions MLP001 AR006460...
Страница 11: ...11 AR006462...
Страница 19: ...19 Connection of the transmitter with the two current outputs passive...
Страница 21: ...21...
Страница 73: ...73 Appendix 2 Output and power supply circuits...
Страница 74: ...74 Appendix 3 Input circuits...