92
The input items include C.O.(Cardiac output), FiO2 (Fraction of inspired oxygen), PaO2 (oxygen partial pres-
sure), PaCO2 (carbon dioxide partial pressure), Hb (Hemoglobin), CaO2 (oxygen content of arterial blood),
CvO2 (oxygen content of mixed blood), VO2 (oxygen consumption), RQ (Respiratory quotient), ATMP (Atmos-
pheric pressure), Height and Weight.
Refer to the following table for the detailed definition, unit and adjustable range of the input items.
Oxygenation calculation function can perform calculation for 14 parameters (i.e. Output items): BSA (body
surface area), VO2 calc (oxygen consumption), C(a- v)O
2
(oxygen content difference between artery and vein),
O
2
ER (oxygen extraction rate), DO
2
(oxygen delivery), PAO
2
(oxygen pressure in alveolar ), AaDO2 (oxygen
pressure difference between alveolar and artery), CcO
2
(pulmonary capillary oxygen content), Qs/Qt (intra-pul-
monary shunt rate), C.O.calc (calculated cardiac output), PaO
2
/ FiO
2
(oxygenation index), AaO
2
/ PaO
2
(Ratio
of alveolar-arterial oxygen partial pressure difference to oxygen partial pressure), DO
2
l (oxygen delivery index)
and VO
2
l (oxygen consumption index).
Refer to the following table for the detailed definition, unit and adjustable range of the output items.
Input item
Unit
Definition
Adjustable range
C.O
L/min
Cardiac output
0.1—20.0
FiO
2
%
Fraction of inspired oxygen
18%--100%
PaO
2
mmHg
Oxygen partial pressure
10—800
PaCO
2
mmHg
Arterial carbon dioxide tension
0—200
Hb
g/L
Hemoglobin
50—200
CaO
2
ml/L
Oxygen content of arterial blood
10—400
CvO
2
ml/L
Oxygen content of mixed blood
10—400
VO
2
ml/min
Oxygen consumption
50—1000
RQ
——
Respiratory quotient
0.1—1.5
ATMP
mmHg
Atmospheric pressure
300—1200
H
cm
Height
20—300
W
kg
Weight
1.0—250.0
Parameters
Definition
Calculation formula
Unit
Range
BSA
Body surface area
W
0.425
*H
0.725
*0.007184
m
2
——
VO2 calc
Oxygen consumption
(SaO2-SvO2)*13.4*Hb * C.O.
ml/min
——
C(a- v)O2
Oxygen content difference
between artery and vein
CaO2 – CvO2
ml/L
42~59
O2ER
Oxygen extraction rate
VO2(CaO2-CvO2)*100%
%
24~28
DO2
Oxygen delivery
CaO2 * C.O.
ml/min
950~1150
PAO2
Oxygen pressure in alveolar
(FiO2*100)*(ATMP - 47) – PaCO2 / RQ)
mmHg
——
AaDO2
Oxygen pressure difference
between alveolar and artery
(FiO2*100)*(ATMP - 47) – (PaCO2 / RQ)
– PaO2
mmHg
10~15
CcO2
Pulmonary capillary oxygen
content
(Hb*1.34)+(((FiO2*100)*(ATMP - 47)
– (PaCO2 / RQ))* 0.0031)
ml/L
——
Qs /Qt
Intra-pulmonary shunt rate
{[(Hb*1.34)+(( (FiO2*100)*(ATMP-47)
–(PaCO2/RQ))* 0.0031)] – CaO2) /
{ [(Hb*1.34)+(( (FiO2*100)*(ATMP-47)
– (PaCO2/RQ))* 0.0031)] – CvO2 }*100
%
3.0~5.0
C.O.calc
Calculated cardiac output
VO2 / (CaO2-CvO2)
L/min
0.1~20.0
PaO2 / FiO2
Oxygenation index
PaO2 / (FiO2*100)
mmHg
——
AaO2 / PaO2
Ratio of alveolar-arterial oxygen
partial pressure difference
to oxygen partial pressure
[(FiO2*100)*(ATMP - 47) – (PaCO2 / RQ)
– PaO2 ] / PaO2
——
——
DO2l
Oxygen delivery index
(CaO2 * C.O.)/BSA
ml/min/m
2
——
VO2l
Oxygen consumption index
(CaO2 – CvO2)*C.O. /BSA
ml/min/m
2
——
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...