53
Caution
: STAT mode can only be used for adult.
Notes:
When changing patient, the default NIBP measuring mode is “Manual”.
For “Adult”, if the NIBP measuring mode is set as “STAT”, then the monitor will not save this setting when
you shutdown the monitor. That’s to say, when re-starting the monitor, the NIBP measuring mode for Adult is
“Manual”.
For all patient types, if the NIBP measuring mode is set as “Manual”,or “Auto”, or “Customized Multi-cycle”,
then the monitor will save this setting when you shutdown the monitor.
•
Cycle
: it can be set only in “AUTO” mode. Cycle means the time interval between measurements when the
measuring mode is set to Auto. The Cycle options are: 1min, 2min, 3min, 4min, 5min, 6min, 7min, 8min,
9min, 10min, 15min, 20min, 25min, 30min, 35min, 40min, 45min, 50min, 55min, 1h, 1.5h, 2h, 2.5h, 3h, 3.5h,
4h, 4.5h, 5h, 5.5h, 6h, 6.5h, 7h, 7.5h and 8h. that’s, the range is from 1 minute to 10 minutes, step is 1 min-
ute; for range from 10minutes~1hour, step is 5 minutes; for range from 1 hour~8 hour, step is 0.5hour. When
in AUTO mode and this item is set, the operator must manually press NIBP measure key “
” for the
first time, then the device will start time counting down. The device will start taking the next blood pressure
measurement automatically after finishing counting down.
•
Unit
: the pressure unit. “mmHg” and “kPa” can be selected. 1 kPa =7.5 mmHg.
•
Init
. Pressure: Cuff pressure to be inflated initially. Its range is different depending on patient type.
for adult
: 80, 100, 120, 140, 150, 160, 180, 200 mmHg;
for pediatric
: 80, 100, 120, 140, 160, 180, 200 mmHg;
for neonate
: 60, 70, 80 mmHg.
Note:
if your monitor is configured with SunTech NIBP module, then the initial cuff pressure range for adult
is 120~220mmHg.
•
PARM
: to enter into NIBP (SYS, DIA, MAP,) alarm setting. See Section “
Understanding the Alarm Setting
”.
•
Verification
: to verify the NIBP pneumatic system, is used to verify the pressure accuracy and check the air
leakage of the pneumatic system, which should be conducted by technicians in the specified environment.
It includes “Verification A”, “Verification B” and “Leakage Check”.
�
Verification A & Verification B
: the 2 kinds of verification of the pressure accuracy. Press the relevant but-
ton to start Verification A or B. When completed, it is necessary to stop the verification manually by pressing
“Stop” button. System will also stop the verification when “Exit” button pressed.
�
Leakage Check
: it is used by technicians to perform a leakage inspection for NIBP pneumatic system.
Important:
Verification A:
the monitor will auto inflate to the preset pressure value (depending on the patient type) by
pump, then it close the deflating valve. Comparing the pressure value on device with that on standard pressure
manometer, and verify that whether the pressure accuracy exceeds the rated tolerance range.
The auto inflating pressure value for Adult: >190mmHg (25.3kPa)
The auto inflating pressure value for Pediatric: >160mmHg (21.3kPa)
The auto inflating pressure value for Neonate: >80mmHg (10.7kPa)
Verification B:
the monitor will close the valve, the pressure should be inflated manually. Comparing the pres-
sure value on device with that on standard pressure manometer, and verify that whether the pressure accuracy
exceeds the rated tolerance range.
Note:
if the verification window pops up during verification, you can press “Close” button to exit form it only,
but it is forbidden to exit from by pressing “Display View Key”.
Chapter 10
MONITORING OXYGEN SATURATION (SpO
2
)
10.1 Introduction
The functional oxygen saturation (SpO
2
) - a percentage of the hemoglobin that can transport oxygen, is mon-
itored by this device via a non-invasive optical technique. Based on the principle that oxygenated hemoglobin
(HbO2) and deoxygenated hemoglobin (Hb) have different absorption character in the spectrum range from
red to infrared light, the device measures the amount of oxygenated hemoglobin and pulse rate by measuring
the absorption of selected wavelengths of light. The light generated in the probe passes through the tissue and
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...
Страница 136: ......
Страница 147: ...11 SERIES II 12 13 Front view Rear view Connector and icon Plugin module installing location Left view Right view...
Страница 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Страница 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Страница 267: ...131...
Страница 268: ......