condition, and the icon “
” displays on the upper right corner of the screen at the same time. Alarm
reset is not a toggle operation, pressing this key once or more times only makes alarm reset.
• The description for technical alarm status indicator and physiological alarm indicator
1) Technical alarm status indicator is an auxiliary indicator for “Alarm lamp”. It’s used for indicating whether
the system in technical alarm status or not. When the system in technical alarm status, then the indicator in
blue, otherwise, it stays off.
2) When high priority physiological alarm event occurs, the physiological alarm indicator “
” is flashing
with red color; when middle priority physiological alarm event occurs, the indicator is flashing with yellow
color; when the low priority physiological alarm event occurs, the indicator is in constant yellow light; if there
is no physiological alarm event, the indicator is in constant blue light.
6.3.6 Changing Alarm Volume
To set the alarm sound volume.
Step 1: Select “Alarm”
“Others”.
Step 2: In “Others”, select “Alarm Volume”. Its setting range is “0 ~ 10”, the step is 1. The default is 5. “10” is
the maximum volume.
When the alarm sound is switched off, the monitor will give no audible alarm tones even if a new alarm
occurs. Therefore the user should be very careful about whether to switch off the alarm sound or not.
Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low
level may result in a hazard to the patient. Always keep the patient under close surveillance.
Pausing or switching off alarms may result in a hazard to the patient, please be very careful.
If occurrence of multiple alarm signal at the same time, the monitor will only show the high priority alarm
in the form of audible alarm and alarm lamp. Meanwhile, all alarm signals information including message
descriptions and numerical highlight etc. will be shown respectively and simultaneously on the screen.
It is suggested that the users should not change the alarm volume lower than the factory default setting if
close and constant attention could not be paid to the patient, otherwise the negligence of alarm event might
cause irreversible harm to the patient.
During the alarm silence period, any new alarm event can activate the audible alarm again and the audible
alarm function resumes normal state.
6.4 Understanding the Alarm Setting
Select “Alarm” button on Statusbar, and select parameter (such ECG, SpO
2
, NIBP, TEMP, etc.). You can re-
view and set alarm limits, alarm switches, high/low alarm limit, and alarm level for each parameter, as shown
in below figure.
Alarm setting window
(for full configuration)
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...
Страница 136: ......
Страница 147: ...11 SERIES II 12 13 Front view Rear view Connector and icon Plugin module installing location Left view Right view...
Страница 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Страница 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Страница 267: ...131...
Страница 268: ......