62
4. The pressure transducer must be placed at the same level with the patient’s heart.
5. Check if the correct label name is chosen, see below table.
6. Zero the transducer.
Label
Description
Label
Description
ART
Artery pressure
LAP
Left atrium pressure
CVP
Central vain pressure
RAP
Right atrium pressure
ICP
Intracranial pressure
PA
Pulmonary pressure
AUXP1
Auxiliary pressure 1
AUXP2
Auxiliary pressure 2
Label and description
IBP transducer kit with disposable pressure sensor
12.4 Understanding the IBP Display
IBP Waveform:
• IBPL1: Parameter label, it means that the left inserted plugin module is the IBP1.
• [0 300] : IBP coordinates scale.
Note:
for some monitors, it may only show 2 channels of waveform.
IBP Panel:
IBP Alarm information
• The IBP alarm information will be displayed on the alarm information area.
• Technical alarm: “Probe off” and so on.
• Physiological alarm: “IBPL1 Low SYS “ and so on.
IBP mark and unit
Systolic pressure
Diastolic pressure
Measuring position
Mean arterial pressure
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...
Страница 136: ......
Страница 147: ...11 SERIES II 12 13 Front view Rear view Connector and icon Plugin module installing location Left view Right view...
Страница 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Страница 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Страница 267: ...131...
Страница 268: ......