Do NOT use the IRMA airway adapter with metered dose inhalers or nebulized medications as this may
affect the light transmission of the airway adapter windows.
Replace the airway adapter if rainout/condensation occurs inside the airway adapter.
Use only PhaseIn manufactured IRMA airway adapter.
The IRMA probe is not intended to be in patient contact.
Never sterilize or immerse the IRMA probe in liquid.
The IRMA airway adapter is non-sterile device. Do not autoclave the devices as this will damage it.
Do not apply tension to the probe cable.
Do not operate the IRMA probe outside the specified operating temperature environment.
15.4.2. Instruction for PHASEIN ISA Sidestream Analyzer
ISA sidestream analyzer is intended to be used for real time monitoring data of CO
2
, N
2
O and the anesthetic
agents (AA) Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane during anaesthesia, recovery and
respiration care. It may be used in the operating suite, intensive care unit, patient room and emergency medi-
cine settings for adult, pediatric and infant patients.
It is NOT intended to be used as the only means of monitoring a patient. It shall always be used in combination
with other vital signs monitoring and/or professional human judgments of patient condition. The ISA side-
stream probe is intended to be used by trained and authorized health car processionals only.
1. System Setting
1. Plug the ISA connector into the host device.
2. Plug the Nomoline sampling line into the ISA input port.
3. Connect the gas outlet with the releasing system, or make the gas flow to the patient circuit.
4. Power on the monitor.
5. A green LED indicates that the ISA gas analyzer is ready for use.
6. Do pre-use check following the section part.
2. Pre-use Check
Before plugging the Nomoline sampling line into the respiratory circuit, please do the following operation:
1. Connect the sampling line to the LEGI.
2. Check whether the green light in LEGI is normal.
3. For ISA OR+ and ISA AX+: Check whether the gas (O
2
) reading is correct (21%).
4. Expiring gas towards the sampling line and check that the gas reading (CO
2
) and waveform are displayed
on the monitor.
5. Clogging the sampling line with finger for 10 seconds.
6. Check if the clogging alarm is displayed and red light in LEGI is blinking.
7. Check whether the patient circuit is hermetic.
3. LEGI Status
Color of LED
Information
Steady green light
System OK
Blinking green light
Zeroing in process
Steady blue light
Anesthetic agent present
Blinking red light
Check sampling line
Steady red light
Sensor error
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...
Страница 136: ......
Страница 147: ...11 SERIES II 12 13 Front view Rear view Connector and icon Plugin module installing location Left view Right view...
Страница 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Страница 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Страница 267: ...131...
Страница 268: ......