58
10.7 Nellcor SpO
2
Module (Optional)
• If your monitor is configured with Nellcor SpO
2
Module, then the SpO
2
related setting window is as shown
in below figure.
•
SE Threshold (%)
: set the threshold of oxygen desaturation event (%), the value range is from 1 to 12, step
is 1. The default is 3(%).
•
SatSeconds
: SatSeconds sensitivity setting, to set the max buffering time of activating SpO
2
alarm. Op-
tions are “0”, “10”, “20”, “50” and “100”. If you set SatSeconds as “0”, it means to turn off the SatSeconds
alarm management function.
The over-limit points multiplied over-limit seconds makes SatSeconds.
SpO
2
panel for Nellcor SpO
2
module:
Notes:
1. When inserting finger into the probe cushion, then searching pulse icon “
” appears on the upper side
of the SpO
2
panel. However, once the SpO
2
/PR readings appear, or probe/ finger is off, then icon “
”
will disappear.
2. When interference (caused by e.g. shaking finger) appears, then icon “
” appears on the upper mid-
dle side of the SpO
2
panel. However, once the interference disappears, or probe / finger is off, then icon
“
” will disappear.
SatSeconds icon
and sensitivity setting
“Interference” icon
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...
Страница 136: ......
Страница 147: ...11 SERIES II 12 13 Front view Rear view Connector and icon Plugin module installing location Left view Right view...
Страница 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Страница 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Страница 267: ...131...
Страница 268: ......