10. NIBP accuracy:
Maximum mean difference: ±5 mmHg
Maximum Standard deviation: 8 mmHg
11. Measurement mode: Manual, Auto, STAT, Customized Multi-cycle
12. NIBP Alarm setting range: see Section Alarms
26.5 SpO
2
1. Transducer: dual-wavelength LED
Wavelength: Red light: 663 nm, Infrared light: 890 nm.
(Note: for Series III and Series IV, Red light: 660 nm, Infrared light: 905 nm)
Maximal optical output power: less than 2mW maximum average
2. SpO
2
measuring range: 0%~100%
3. SpO
2
measuring accuracy: Arms* is not greater than 2% for SpO
2
range from 70% to 100%
*NOTE: Arms is the accuracy defined as root-mean-square value of deviation according to ISO 80601-2-61.
4. Alarm range:
High: 1%~100%; Low: 0%~99%
5. Low perfusion performance: the declared accuracy is sustained when the pulse amplitude modulation ratio
is as low as 0.3%
Data averaging and update:
The displayed SpO
2
and Pulse Rate values are the average of data collected within a specific time. The SpO
2
is calculated every second by the data collected in recent 5 seconds, the Pulse Rate is calculated for every
beat. The averaging method depends on the pulse rate value, for pulse rates below 50bpm, the SpO
2
is av-
eraged by 16-second sliding average, the Pulse Rate is averaged by 4-beat sliding average; for pulse rates
between 50bpm and 120bpm, the SpO
2
is averaged by 8-second sliding average, the Pulse Rate is averaged
by 8-beat sliding average; for pulse rates above 120bpm, the SpO
2
is averaged by 4-second sliding average,
the Pulse Rate is averaged by 16-beat sliding average.
The screen display of SpO
2
and Pulse Rate are updated every second with the most recent value, if the signal
is noisy or missing, the display will hold the last value for at most 15 seconds before showing dashes.
The over-limit alarm is triggered once the SpO
2
or Pulse Rate exceeds the preset limits, the alarm signal gen-
eration delay is small (less than 1 second) compared with the alarm condition delay, which is caused by the
data average as described above.
26.6 Pulse Rate
1. Pulse rate measuring range: 0bpm~250bpm
2. Pulse rate measurement accuracy: ±2bpm or ±2%, whichever is greater.
3. Alarm range:
High: (1~300) bpm; Low: (0~299) bpm
4. Pulse rate alarm tolerance: ±2bpm within (30~250) bpm
Note: The pulse rate ACCURACY is tested by an electronic pulse simulator.
26.7 CO
2
1. Technology: Infrared absorption method.
2. Mode of Sampling: Sidestream or Mainstream
3. CO
2
Response Time:
Sidestream: <3seconds (including transport time and rise time).
Mainstream: <60ms (rise time)
4. Warm-up Time: Not less than two minutes
5. CO
2
measurement range: 0~150mmHg
6. CO
2
Accuracy: 0~40mmHg
±2mmHg
41~70mmHg
±5% of reading
71~100mmHg
±8% of reading
101~150mmHg
±10% of reading
7. Alarm range:
High limit: 0.1mmHg ~150mmHg
Low limit: 0mmHg ~149.9mmHg
*NOTE: Gas temperature at 25°C for Sidestream;
Gas temperature at 35°C for Mainstream
8. Flow rate: 50ml/min ±10 ml/min (Sidestream)
Содержание 35145
Страница 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Страница 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Страница 135: ...135...
Страница 136: ......
Страница 147: ...11 SERIES II 12 13 Front view Rear view Connector and icon Plugin module installing location Left view Right view...
Страница 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Страница 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Страница 267: ...131...
Страница 268: ......