62
•
Operational Reference
Genie Monochrome Series-GigE Vision Camera
Using Flat Field Correction
From the CamExpert menu bar enable Flat Field correction (
Pre-Processing • Flat Field Correction • Hardware
).
Now when doing a live grab or snap, the incoming image is corrected by the current flat field calibration data for
each pixel.
Use the menu function
Tools • Flat Field Correction • Load
to load in a flat field correction image from
previously saved calibration data. CamExpert allows saving and loading calibration data for all cameras used with
the imaging system.
Flat Field Correction Control via Sapera LT or GigE Vision Compliant
Applications
Sapera
Genie Sapera parameters for Flat Field Correction:
FlatFieldEnable
= Enable Flat Field Correction {False=0, True=1}
FlatFieldFormat
= {RO: mono 8, mono 9, mono 10, RGB}
FlatFieldWidth
= read buffer width
FlatFieldHeight
= read buffer height
FlatFieldGainMin
= read minimum gain value
FlatFieldGainMax
= read maximum gain value
FlatFieldOffsetMin
= read minimum offset value
FlatFieldOffsetMax
= read maximum offset value
FlatFieldGainBuffer
= Flat Field correction gain data buffer
FlatFieldOffsetBuffer
= Flat Field correction offset data buffer
+ Class constructor: SapFlatField::SapFlatField
In CamExpert, these features are available from the “Pre-Processing” menu.
XML features for Flat Field Correction:
Not supported
Содержание M1024
Страница 88: ...86 Network Overview Tools Genie Monochrome Series GigE Vision Camera...
Страница 111: ...Genie Monochrome Series GigE Vision Camera Technical Specifications 109 EC FCC Declaration of Conformity...
Страница 112: ......
Страница 126: ...124 Troubleshooting Genie Monochrome Series GigE Vision Camera...
Страница 132: ...130 Glossary of Terms Genie Monochrome Series GigE Vision Camera...