Genie Monochrome Series-GigE Vision Camera
Operational Reference
•
61
CamExpert Flat Field Calibration Menu
•
Click on the
Advanced Setting
button to change the default number of frames averaged for each calibration
step. The default value is 10 frames.
•
Setup the camera to capture a uniform black image. Black paper with no illumination and the camera lens’ iris
closed to minimum can provide such a black image.
•
Click on
Acquire Black Image
. The flat field calibration tool will grab video frames, analyze the pixel gray
level spread, and present the statistics. The desired black reference image should have pixel values less then 20.
If the results are acceptable, accept the image as the black reference.
•
Setup the camera to acquire a uniform white image (but not saturated white). Even illumination on white paper
can be used, with a gray level of 128 minimum. It is preferable to prepare for the white level calibration step
before starting the calibration procedure.
•
Click on
Acquire White Image
. The flat field demo will grab video frames, analyze the pixel gray level
spread, and present the statistics. The captured gray level for all pixels should be greater than 128. If acceptable
accept the image as the white reference.
•
Click on
Save
. The flat field correction data is saved as a TIF image with a file name of your choice
(suggestions are the camera name and its serial number).
Содержание M1024
Страница 88: ...86 Network Overview Tools Genie Monochrome Series GigE Vision Camera...
Страница 111: ...Genie Monochrome Series GigE Vision Camera Technical Specifications 109 EC FCC Declaration of Conformity...
Страница 112: ......
Страница 126: ...124 Troubleshooting Genie Monochrome Series GigE Vision Camera...
Страница 132: ...130 Glossary of Terms Genie Monochrome Series GigE Vision Camera...