The unit can be used by 1 or 2 twin foot
‐
switches (See the paragraph USE OF THE FOOT
‐
SWITCHES”). The
unit, for this endoscopy, is normally used by the 34-9333 model which is provided with a foot selector to activate
either the monopolar currents or the bipolar currents.
Connect the 34-9333 to the 1 socket.
To deliver the monopolar currents push, for an instant, the foot selector. The unit signals the selection in
three ways: MONO word by digital speech, MONO 1 image ON for 2 seconds on the touch
‐
screen. (1a
–
MONO light) of the PEDAL
‐
1 socket ON. To avoid mistakes of use, block the foot selector of the 34-
9333, by the specific protective device.
CONNECTION OF THE ACCESSORIES
NEUTRAL ELECTRODE.
To use the neutral electrode follow all the related warnings of the Par. BASIC WARNINGS. For the
operation of its control circuit see the Par. CONTROL CIRCUIT OF THE NEUTRAL ELECTRODE. The
electrode, during pauses in use, can be disconnected. See the Par. STANDBY. Connect the electrode to
the 7 socket (either7a or 7b according to the connector of the cable).
RESECTOSCOPE OR OTHER INSTRUMENT.
Connect to the 5 socket (Usable when the 5a light is ON) the resectoscope / other instrument. (The plug
of the cable
is the “Ø 6mm” type).
INDICATIONS FOR THE SUITABLE CURRENTS
TURP
ELECTRODE
CURRENTS AND STARTING SETTING
PURE CUT
Wire loop
PURE CUT: 110/120 or AUTO PURE MICRO: Effect 8
COAGULATING CUT
Wire loop
BLEND CUT 1: 110/120 or AUTO BLEND MICRO: Effect 8
PURE CUT
Ribbon loop.
PURE CUT: 130/140 or AUTO PURE MICRO: Effect 9
COAGULATING CUT
Ribbon loop.
BLEND CUT 1: 110/120 or AUTO BLEND MICRO: Effect 9
COAGULATION
Wire loop/Ribbon loop
FULG FORCED COAG: 70/80 or SPRAY COAG: 60/70
COAGULATION
Ball or Roll
FULG FORCED COAG: 80/90 or SPRAY COAG: 70/80
TURV / GYNAECOLOGY
PURE CUT
Wire loop
PURE CUT: 100/110 or AUTO PURE MICRO: Effect 7/8
COAGULATING CUT
Wire loop
BLEND CUT 1: 110/120 or AUTO BLEND MICRO: Effect 8
PURE CUT
Ribbon loop.
BLEND CUT 1: 100/110 or AUTO BLEND MICRO: Effect 8
COAGULATING CUT
Ribbon loop.
BLEND CUT 1: 120/130 or AUTO BLEND MICRO: Effect 9
COAGULATION
Wire loop/Ribbon loop
FULG FORCED COAG: 70/80 or SPRAY COAG: 60/70
COAGULATION
Ball or Roll
FULG FORCED COAG: 80/90 or SPRAY COAG: 70/80
VAPORIZATION
Ball or Roll (3mm)
PURE CUT: 140/150
VAPORIZATION
Ball or Roll (5mm)
PURE CUT: 190/200
PURE CUT
Knife “L”
PURE CUT: 60/70 or AUTO PURE MICRO: Effect 5
COAGULATING CUT
Knife “L”
BLEND CUT 1: 60/70 or AUTO BLEND MICRO: Effect 5
ARTHROSCOPY
Reduce the previous setting to about the 50%
DELIVERY OF THE MONOPOLAR CURRENTS
(MONOPOLAR CUT / COAGULATING CUT) Set the MONO mode and push the yellow pedal. The
delivery is signalled by an acoustic signal (low, adjustable in the GENERAL MENU) and by the big
image on the touch screen.
(MONOPOLAR COAGULATION) Set the MONO mode and push the blue pedal. The delivery is
signalled by an acoustic signal (acute, adjustable in the GENERAL MENU) and by the big image on the
touch screen
THE ALARM SIGNALS DURING USE
During the use, in case of breakages, problems or mistakes of use, the self
‐
check system stops the power
delivery and it informs users by acoustic and visual signals. In case of immediately solvable problems, users can
intervene to eliminate causes. For other signals users can switch the unit OFF/ON to verify them and ask for
technical assistance.
THE MOST COMMON SIGNALS