The lights of the sockets switch ON, according to the selected use, to highlight the usable socket.
A
Touch-screen
B
PS Keys to set the power of currents
1, 2
Sockets to connect the twin foot-switches
3, 5
Sockets to connect the monopolar accessories
4
Socket to connect either a monopolar electrode
7
Socket to connect the neutral electrode
8
Sockets to connect the bipolar accessories
C
Mains switch
E
Plug for the equipotential connection
F
Socket for the supply cord with fuse block
D, G, H
Not used
Alternating current (on the data label)
Earth protection (inside the unit)
Apparatus of Class I Type CF, protected against the defibrillator effects (a CF type unit guarantees the
highest safety level against direct and indirect contacts, notably for the allowable leakage currents). The
applied part type F (floating) is protected from the earth at high and low frequencies. This kind of unit is
especially indicated for direct heart application.
GEISTER MEDIZINTECHNIK GMBH guarantees safety, reliability and performances of this unit only if
installation, recalibrations and repairs are carried out, using original spare parts, by personnel
authorized by GEISTER and if the unit is used in compliance with given instructions in an area that
meets all the applicable IEC or CEI requirements.
GEISTER is at disposal to supply, if requested, the electric diagrams and any further information needed.
This manual must be kept where the unit is employed. Please read this entire manual carefully to become
familiar with each of the controls and features before making any attempt to use the equipment clinically and ask
it again in case missing.
If any questions arise regarding the information contained in this manual according to specific needs, please
contact the Manufacturer, directly or through the local distributor, before using the unit. In accordance with the
requirements of the European Directive for medical devices 93/42 CEE and with the procedures of Company
Quality System for the after-sale control of the production, the users are pleased to inform the Manufacturer
about every, even little, problem of this unit.
PEDAL
‐
1 and PEDAL
– 2 sockets
PEDAL
‐
1 socket: Connection of the 34-9333 standard foot
‐
switch provided with a foot
‐
selector to
activate either the monopolar currents or the bipolar currents. The model 34-9332 (Identical but larger for
the endoscopy) is available on request.
PEDAL
‐
2 socket: Connection of the 34-9331 foot
‐
switch to activate the bipolar currents only. When
connecting both foot
‐
switches, the 34-9333 activates the monopolar currents and the 34-9331 activates
the bipolar currents.