TURVis or GYNAECOLOGY with normal resectoscopes
CUT
Wire loop / Ribbon loop.
SALINE URO
‐
GYN CUT: Effect 4/7
(Continuous delivery or Pulsed delivery)
COAGULATION
Wire loop / Ribbon loop.
SOFT MICRO BICOAG: 125
COAGULATION
Ball / Roll
SOFT MICRO BICOAG: 125
VAPORIZATION
Ball / Roll
SALINE URO
‐
GYN CUT: Effect 8 (Continuous delivery)
CUT
“L” Knife
SALINE URO
‐
GYN CUT: Effect 3/5
(Continuous delivery or Pulsed delivery)
TURVis or GYNAECOLOGY with mini resectoscopes
CUT
Wire loop / Ribbon loop.
SALINE URO
‐
GYN CUT: Effect 3/5
(Continuous delivery or Pulsed delivery)
COAGULATION
Wire loop / Ribbon loop.
SOFT MICRO BICOAG: 70/120
FORCED MACRO BICOAG: 70/120
COAGULATION
Ball / Roll
SOFT MICRO BICOAG: 110/120
FORCED MACRO BICOAG: 110/120
VAPORIZATION
Ball / Roll
SALINE URO
‐
GYN CUT: Effect 8 (Continuous delivery)
CUT
“L” Knife
SALINE URO
‐
GYN CUT: Effect 3/5
(Continuous delivery or Pulsed delivery)
ARTHROSCOPY
Decrease the starting setting to about the 50% of the previous ones
When using the SALINE ARTHRO CUT current: Effect 4/5
DELIVERY OF THE BIPOLAR CURRENTS
(CUT WITH CONTINUOUS DELIVERY) Push the P key to switch OFF it and push the yellow pedal. The
delivery is signalled by an acoustic signal (low, adjustable in the GENERAL MENU) and by the big
image on the touch screen.
(CUT WITH PULSED DELIVERY) Push the P key to switch ON it and push the yellow pedal. The
delivery is signalled by an acoustic signal (low, adjustable in the GENERAL MENU) and by the big
image on the touch screen.
(COAGULATION) Push the blue pedal. The delivery is signalled by an acoustic signal (acute, adjustable
in the GENERAL MENU) and by the big image on the touch screen
Continuous or pulsed delivery of the currents: Saline Uro
‐
Gyn Cut and Saline Arthro Cut. Use MS keys that switch
ON/OFF by pushing them.
Keys OFF: By pushing the foot
‐
switch, the current delivery is continuous.
C1 key ON: By pushing the foot
‐
switch, the unit delivers only a single pulse of current.
CCS key ON: By pushing the foot
‐
switch, the current delivery is continuous, but pulsed (Slow
modulation).
CCF keyON: By pushing the foot
‐
switch, the current delivery is continuous, but pulsed (Fast modulation).
THE ALARM SIGNALS DURING USE
During the use, in case of breakages, problems or mistakes of use, the self
‐
check system stops the power
delivery and it informs users by acoustic and visual signals. In case of immediately solvable problems, users can
intervene to eliminate causes. For other signals users can switch the unit OFF/ON to verify them and ask for
technical assistance.
THE MOST COMMON SIGNALS
See the TABLE I of the paragraph SELF
‐
CHECK SYSTEM where the complete operation of the self
‐
check
system is detailed.
Use of the bipolar currents only (MEMORIES FROM 93 TO 99)
These memories allow the use of all the bipolar currents by one or two foot
‐
switched instruments. The AUTO
MICRO SOFT and AUTO SEALING currents are provided with automatic “impedance sensing” start/stop.
WARNINGS AND PRECAUTIONS FOR THE OPERATION
To perform all operations, carefully apply all the warnings of the paragraph BASIC WARNINGS!
SETTING OF THE CURRENTS
To set, memorize and change the currents. Follow the instructions of the paragraph FIRST SETTING.
To choose the best currents and the starting powers for the specific need. See, in this section, the
paragraph INDICATIONS FOR THE SUITABLE CURRENTS.