To change the MONOPOLAR PURE CUT current,
do the following:
Push the PS keys on the panel. On the
area (1) the watts change.
To change the AUTO PURE MICRO current, do
the following:
Push the PS keys on the panel. On the
area (1), the following changes: The
effects (from 1 to 10) and the PWR (max.
power deliverable with the effect).
Change these currents with care! When
changing the effect, the power changes
notably!
Use, in open or laparoscopic surgery, of the monopolar and bipolar currents
(MEMORIES FROM 1 TO 64)
These memories allow the use of the following currents:
All the monopolar and bipolar currents.
The monopolar currents are usable by:
One hand
‐
switched or foot
‐
switched handle / instrument.
One foot
‐
switched handle / instrument and one hand
‐
switched handle.
Two hand
‐
switched handles.
The bipolar currents are always usable by:
One or two foot
‐
switched instruments. Two currents, AUTO MICRO SOFT and AUTO SEALING, are
provided with automatic “impedance sensing” start/stop.
WARNINGS AND PRECAUTIONS FOR THE OPERATION
To perform all operations, carefully apply all the warnings of the paragraph BASIC WARNINGS!
SETTING OF THE CURRENTS
To set, memorize and change the currents. Follow the instructions of the paragraph FIRST SETTING.
To choose the best currents and the starting powers for the specific need. See, in this section, the
paragraph INDICATIONS FOR THE SUITABLE CURRENTS.
CONNECTION AND USE OF THE FOOT
‐
SWITCHES
The unit can be used by 1 or 2 twin foot
‐
switches.
The standard 34-9333 model (On request it is also available the large 34-9332 model for endoscopy)
provided with a foot selector to activate either the monopolar currents or the bipolar currents.
The 34-9331 model, on request, which activates the bipolar currents only. See, in the paragraph USE
OF THE FOOT
‐
SWITCHES, the complete use of the foot
‐
switches.
To connect the 34-9333, use the 1 socket, to connect the 34-9331 use the 2 socket.
Use of the 34-9333 foot
‐
switch.
To deliver the monopolar currents push, for an instant, the foot selector.
The unit signals the selection in three ways:
MONO
word by digital speech, MONO 1 image ON for 2 seconds on the touch
‐
screen. (1a
– MONO
light) of the PEDAL
‐
1 socket ON.
To deliver the bipolar currents push, for an instant, the foot selector.
The unit signals the selection in three ways: