DIAGRAMS OF THE BIPOLAR CURRENTS
Values measured within 3 seconds, by excluding the transients lower than 1 second, as established in IEC
60601
‐
2
‐
2. With regard to the powers, we stress the following:
The allowed tolerance is 20% for the powers higher than the 10% of the maximum power of the
current.
The measurements have to consider the tolerance of the devices used to perform the measurement.