2.1
SECTION: SOUND
By rotating the control when the BEEP is ON, users can change the level of the sounds which signal
both the delivery of currents and the use of keys. All acoustic alarm signals are fixed, not adjustable.
2.2
SECTION: BEEP ON/OFF
By pushing the key, users can set the acoustic signal of the keys on the panel.
2.3
SECTION: NEUTRAL PLATE SOCKET SELECTION
By using the key, users can select the socket to connect the neutral electrode:
Socket to connect a cable with European, standard (jack Ø 6,35 mm) connector.
Socket to connect a cable with American, Valleylab type, connectors.
See the paragraph CONTROL CIRCUIT OF THE NEUTRAL ELECTRODE.
2.4
SECTION: ACTIVATION PEDAL 1 SELECTION
By pushing the key, users can manually set the functioning mode of the standard, main, twin foot
‐
switch 34-9333. See the paragraph CONNECTION AND USE OF FOOT
‐
SWITCHES.
2.5
SECTION: BIPOLAR OUTPUT SELECTION
By pushing the key, users can select the sockets to connect the bipolar instruments:
Only the BIPOLAR
‐
2 socket / Both sockets: BIPOLAR
‐
1 and BIPOLAR
‐
2.
2.6
SECTION: AUTO BIP START DELAY
By rotating the control, users can set the delay (from 0.1 to 5 seconds) of activation of the bipolar
currents with
automatic “impedance sensing” start/stop:
AUTO SOFT MICRO BICOAG for coagulation / AUTO SEALING for vessel sealing.
2.7
SECTION: LCD color
By pushing the key, users can set the colour (green/red) of the numbers on the touch screen.
LANGUAGE OF THE TOUCH
‐
SCREEN
To set the language, push the display 2 of the upper part of the touch
‐
screen
until opening the specific page of the menu and push the button of the
language.
SELF
‐
CHECK
THE UNIT IS PROVIDED WITH A SELF
‐
CHECK SYSTEM OF THE FUNCTIONING
The system, at the switching ON, performs a complete self
‐
check which ends only if the unit operates perfectly.
The unit signals this phase, by specific codes on the touch screen. The system, during use, performs the
following:
It always checks both the functioning and the delivered powers.
It repeats, every about 20 minutes the complete self
‐
test without disturbing the normal use.
It also checks the mistakes of use (For example the not correct use of an activation switch).
In all cases the system, if it detects both breakages and mistakes of use, blocks the the delivery of powers by
signalling the problem with specific alarms and error codes. In case of the signals detailed in the TABLE I, users
can eliminate the cause. In case of other signals, users can switch the unit OFF/ON to verify the alarm and, if it is
confirmed, ask for the Technical assistance.
MEMORIZATION OF THE DETECTED ERROR CODES.