Dateiname mit Revisionsstand:
erstellt am:
erstellt von:
geprüft von:
freigegeben von:
Seite
ESU-X 500TT_EN B
01.06.15
cgro
cg
cgro
26 von 45
BIP
word by digital speech, BIP 1 image ON for 2 seconds on the touch
‐
screen. (1b
– BIP light) of the
PEDAL
‐
1 socket ON.
Use of the foot
‐
switches 34-9333 and 34-9331.
The monopolar currents are activated by the 34-9333, the bipolar ones by the 34-9331. When
connecting both foot
‐
switches, the unit selects this mode automatically.
CONNECTION OF THE ACCESSORIES
NEUTRAL ELECTRODE.
To use the neutral electrode follow all the related warnings of the Par. BASIC WARNINGS. For the
operation of its control circuit see the Par. CONTROL CIRCUIT OF THE NEUTRAL ELECTRODE. bThe
electrode, during pauses in use, can be disconnected. See the Par. STANDBY.
Connect the electrode to the 7 socket (either7a or 7b according to the connector of the cable).
MONOPOLAR INSTRUMENTS - The following accessories are usable:
One hand
‐
switched or one foot
‐
switched handle / instrument.
One hand
‐
switched handle /instrument and one foot
‐
switched handle /instrument.
Two hand
‐
switched handles.
Connect to the 4 socket (Usable when the 4a light is ON) the hand
‐
switched handle /instrument. (The
plug of the
cable is the”3 pins” type).
Connect to the 5 socket (Usable when the 5a light is ON) the foot
‐
switched handle /instrument. (The plug
of the cable is the “Ø 6mm” type).
To connect two hand
‐
switched handles, use the 3 and 4 sockets (Usable when the 3a and 3b lights are
ON) (The plug of the cable is the”3 pins” type).
BIPOLAR INSTRUMENTS.
The unit allows the connection and the use of one (BIP
‐
2) or two instruments(BIP 1+2) by using the key
of the section 2.5 of the GENERAL MENU.
By using one instrument: All currents are delivered through it.
By using two instruments: The currents for cut are delivered through by first instrument. The currents for
coagulation/vessel sealing are delivered by the second instrument.
Connect to the 8b socket (Usable when the 8b light is ON) one instrument.
Connect to the 8a and 8b sockets (Usable when both lights, 8a and 8b, are ON) two instruments. (The
connector is “2 x Ø 6mm plugs” type and it does not require a specific polarity of connection).
INDICATIONS FOR THE SUITABLE CURRENTS
CURRENTS FOR MONOPOLAR CUT OR COAGULATING CUT. PURE CUT
– Cut without coagulating effect. It
is suitable for the following:
To cut, without coagulating effect, all tissues in all operating fields when using (starting from 50/60)
either needle or knife electrodes for Open surgery or the hook / the side of a dissector for Laparoscopy.
To reduce the smoke in Laparoscopy, the intermittent activation, while grazing the tissues with the
electrode, is advisable.
BLEND CUT
‐
1
‐
Cut with soft coagulating effect. It is suitable for the following:
To cut, with soft coagulating effect, all tissues in all operating fields when using (starting from 50/60)
either needle or knife electrodes for Open surgery or the hook / the side of a dissector for Laparoscopy.
To reduce the smoke in Laparoscopy, the intermittent activation, while grazing the tissues with the
electrode, is advisable.
BLEND CUT 2
– It includes cut (2%) and coagulation (98%) and it allows the obtaining all the effects of the
FULG FORCED COAG, it means efficacious coagulation and cut with strong coagulating effect.
It is suitable for the following:
To dissect, with very strong “SPRAY” coagulating effect, all tissues in all operating fields when using
(starting from 50/60) either needle or knife electrodes for Open surgery or the hook / the side of a
dissector for Laparoscopy. To reduce the smoke in Laparoscopy, the intermittent activation, while
grazing the tissues with the electrode, is advisable.
To coagulate tissues only, by using the above accessories.
PURE CUT PULSED
‐
Current for pulsed cut without coagulating effect. It is suitable for the following:
For the same use of the PURE CUT current (starting from 50/60) while grazing the tissues with the
electrode. It is very useful for the Laparoscopy, to reduce the smoke.
BLEND CUT PULSED
‐
Current for pulsed cut with coagulating effect. It is suitable for the following:
For the same uses of the PURE CUT PULSED current (starting from 50/60).