Dateiname mit Revisionsstand:
erstellt am:
erstellt von:
geprüft von:
freigegeben von:
Seite
ESU-X 500TT_EN B
01.06.15
cgro
cg
cgro
12 von 45
MONOPOLAR
– 2 socket
When the 4a HAND
‐
SWITCHED light is ON. It allows the use of a hand
‐
switched handle or instrument.
The socket allows the use of accessories with the connector “3 pins
‐
international type”.
BIPOLAR
– 1 and BIPOLAR
‐
2 sockets
When the 8b light is ON : only the BIPOLAR
‐
2 socket is usable. All the currents, for cut and coagulation
or vessel sealing, are delivered through it.
When 8b and 8a lights are ON : both sockets, BIPOLAR
‐
1 e BIPOLAR
‐
2, are usable. The currents for
cut are delivered through the first socket, the currents for coagulation or vessel sealing are delivered
through the second socket.
The sockets allow the use of accessories with 2 plugs Ø 4 mm connector
To select them, see the GENERAL MENU, section “bipolar output selection”.
To use accessories with cables having connectors different from the standard one, use the following
adaptors:
RD/BF1: To use cables with connector ERBE, STORZ type.
RD/BF2: To use cables with connector MARTIN type.