The unit allows the fast selection of a program before use.
The unit includes 15 memories for the bipolar use only, without the neutral electrode.
The unit includes some programs with a specific presetting to facilitate their use.
All settings of a program can be changed whenever users like
The memories (from 1
‐
to
‐
64) are for open or laparoscopic surgery by delivering: All normal, monopolar
and bipolar, currents.
The memories (from 65
‐
to
‐
74) are for the flexible endoscopy by delivering: All normal, monopolar and
bipolar, currents.
The memories (from 75 to 84) are for the, under liquid, monopolar endoscopy in UROLOGY,
GYNAECOLOGY AND ARTHROSCOPY.
The memories (from 85 to 99) are for the bipolar use only.
The memories from 85 to 89 are for the endoscopy in saline (UROLOGY and GYNAECOLOGY).
The memories from 89 to 92 are for the arthroscopy.
The memories from 93 to 99 are for open or laparoscopic surgery.
All the controls to use memories are in the same area of the touch
‐
screen.
To select a memory, see the Par: SELECTION OF A MEMORY.
To perform the first setting and the memorization of a program, see the Par: FIRST SETTING
To change the setting memorized in a program, see the Par: ADJUSTMENTS DURING THE USE.
To either cancel or memorize the changes of settings, see the Par: CANCELLATION OR
MEMORIZATION OF THE CHANGES OF THE SETTING.
SELECTION OF A MEMORY
Select the memory, either to program it or to select it for the use, as follows:
By scrolling all memory numbers. Push any of the two 3 keys.
By selecting the memory number. Push the 2 area (The screen 1 appears). In the screen write the
memory number. Push the exit key.
Users can select a new memory only if the memory number is not blinking! See the following paragraph.
BLINKING OF A MEMORY NUMBER
The memory number blinks when users either perform the first setting of a memory or, during use, they change
the already memorized settings.
The blinking does not signal problems and the unit goes on operating normally.
If the number is blinking at the switching OFF, it blinks again at the following switching ON and the unit
resets the, previously used, settings.
The unit doesn’t reset the selection of two bipolar currents: AUTO SOFT MICRO BICOAG, AUTO
SEALING. These currents are provided with automatic start/stop and IEC 60601
‐
2
‐
2 standards for the
safety of HF surgical units establish that users must select currents with automatic start/stop only
intentionally.