See the TABLE I of the paragraph SELF
‐
CHECK SYSTEM where the complete operation of the self
‐
check
system is detailed.
Use for the, in saline, endoscopy and arthroscopy of the bipolar currents
(MEMORIES FROM 85 TO 92)
These memories allow the delivery of the only currents suitable for this use:
The SALINE current for cut and the SOFT MICRO or FORCED MACRO currents for coagulation which usable by
a foot
‐
switched bipolar accessory: a resectoscope, another instrument (for example a flexible, 5Cr, needle for
gynaecology) and an instrument for arthroscopy. In the memories, from 85 to 88, for UROLOGY (TURPis,
TURVis) or GYNAECOLOGY, the suitable currents are already preset:
SALINE URO
‐
GYN CUT current for cut with continuous or pulsed delivery.
SOFT MICRO BICOAG or FORCED MACRO BICOAG currents for coagulation.
In the memories, from 89 to 92, for ARTHROSCOPY, the suitable, currents are already preset:
SALINE ARTHRO CUT current for cut with continuous or pulsed delivery.
SOFT MICRO BICOAG or FORCED MACRO BICOAG currents for coagulation.
WARNINGS AND PRECAUTIONS FOR THE OPERATION
To perform all operations, carefully apply all the warnings of the paragraph BASIC WARNINGS!
SETTING OF THE CURRENTS
To set, memorize and change the currents. Follow the instructions of the paragraph FIRST SETTING.
To choose the best currents and the starting powers for the specific need. See, in this section, the
paragraph INDICATIONS FOR THE SUITABLE CURRENTS.
CONNECTION AND USE OF THE FOOT
‐
SWITCHES
The unit can be used by 1 or 2 twin foot
‐
switches.
The standard
34-9333
model (On request it is also available the large
34-9332
model for endoscopy)
provided with a foot selector to activate either the monopolar currents or the bipolar currents.
The
34-9331
model, on request, which activates the bipolar currents only.
To connect the
34-9333,
use the
1
socket, to connect the
34-9331
use the
2
socket.
When selecting the programs for the bipolar use only (from 85 to 99) all foot
‐
switches activate the bipolar currents
only (The foot
‐
selector of the models 34-9333 and 34-9332
doesn’t operate).
CONNECTION OF THE ACCESSORIES
NEUTRAL ELECTRODE. The electrode is not required and the control circuit is blocked. The alarm lights are ON,
but for information only. RESECTOSCOPE OR OTHER INSTRUMENT FOR ENDOSCOPY OR
ARTHROSCOPY.
In the GENERAL MENU, point 2.5, select the 8b socket by using the related key (the 8b light is ON) to
connect the cable of the electrode.
INDICATIONS FOR THE SUITABLE CURRENTS
To obtain the intended functioning use saline solutions (0,9 of sodium chloride) and not solutions with different
salts (I.E. The Ringer solution which includes potassium chloride).
To obtain a cut very precise and perfectly controlled, for example while using small resectoscopes in
gynaecology, select the pulsed delivery.
To obtain the best result, when delivering the current for cut, slightly warm the saline solution (warm it by the
heating device for the bags or for their flow).
To coagulate, touch vessels until the coagulating effect reaches the surrounding tissues.
To avoid problems when using a resector, mainly when sterilizing it by liquids, dry very careful all its parts and the
connecting cable since even the only humidity can cause two risks: Either the resector operates in unstable way
(It doesn’t operate since the current doesn’t reach the operative end of the electrode) or the insulation of the
instrument could be damaged.
TURPis in
UROLOGY
ELECTRODE
CURRENTS AND STARTING SETTING
CUT
Wire loop / Ribbon loop.
SALINE URO
‐
GYN CUT: Effect 6/9 (Continuous delivery)
COAGULATION
Wire loop / Ribbon loop.
SOFT MICRO BICOAG: 125
FORCED MACRO BICOAG: 125
COAGULATION
Ball / Roll
SOFT MICRO BICOAG: 125
FORCED MACRO BICOAG: 125