Memorization of the setting
When arranging one program only.
By the 4 key, memorize the setting of the memory. The memory number stops blinking.
When arranging more programs.
By the 4 key, memorize the setting of the first memory. The memory number stops blinking.
Select a new memory as it is detailed in the paragraph SELECTION OF A MEMORY.
Perform its setting by repeating the same way.
By the 4 key, memorize the setting of the new memory. The unit allows the memorization of 99
memories! Remember that the system, if the memory number is blinking, does not allow the selection of
a new memory!
Return to the normal use
Connect the neutral electrode when selecting the memories for the monopolar use (1
– 84)!
Select the memory to use as the paragraph SELECTION OF A MEMORY details.
Push the standby key (It switches OFF) to use the unit.
ADJUSTMENTS DURING THE USE
The setting of a program can be changed, during the use, in three ways:
While not activating the power delivery, all the setting of a program can be changed completely.
While not activating the power delivery, the power of the selected currents can be changed.
While activating the power delivery, the power of the delivered current can be changed.
Blinking of the memory number.
When users change a memory, its number blinks, but the blinking does not signal problems and the unit goes on
operating normally. The memory number stops blinking when users either cancel or memorize the changes of
setting. See the paragraph BLINKING OF A MEMORY NUMBER.
Change of the setting of a program while not activating the power delivery
To change the general modes of functioning do the following:
By pushing the menu key of the touch
‐
screen, select the GENERAL MENU and perform the setting as it
is detailed in the paragraph GENERAL MENU.
To change the setting of the monopolar or bipolar currents do the following:
Follow the instructions detailed in the paragraph FIRST SETTING, point 4): Setting, in all memories, of
the monopolar and bipolar currents.
Change of the power of the selected currents while not activating the delivery
For example, in the image the following currents are selected:
Two monopolar currents: PURE CUT and FULG FORCED COAG.
Two bipolar currents: STANDARD BICUT and SOFT MICRO BICOAG.
To change the power of the currents, do the following:
Push the 1 area of the current to change (Its frame switches ON)
By using the PS keys on the panel set the powers
Change of the power of delivered current while activating the power delivery
When users activate a current, on the touch
‐
screen the large image of the current appears. Two examples are
here under included:
MONOPOLAR PURE CUT with ADC self
–control (Indication of the power = Watts).
MICRO PURE CUT with APC self
–control (Indication of the power = Effects).