Dateiname mit Revisionsstand:
erstellt am:
erstellt von:
geprüft von:
freigegeben von:
Seite
ESU-X 500TT_EN B
01.06.15
cgro
cg
cgro
27 von 45
AUTO PURE MICRO
‐
Current for delicate cut without coagulating effect. It is suitable for the following:
For the same uses of the PURE CUT current (starting from the effect nr. 6) when users prefer a more
delicate effect.
AUTO BLEND CUT MICRO
‐
Current for delicate cut with coagulating effect. It is suitable for the following:
For the same uses of the BLEND CUT 1 current (starting from the 6 effect) when users prefer a more
delicate effect.
AUTO PAPILLO PURE CUT, AUTO POLIPO BLEND CUT and AUTO ENDO CUT currents are not suitable for
this use CURRENTS FOR MONOPOLAR COAGULATION. FULG FORCED COAG
‐
Coagulation with strong
sparking. It is suitable for the following:
To coagulate (starting from 50/70), with strong superficial and deep effect, all tissues in all operating
fields, both by grazing the tissues with the active electrode and by touching them with surgical forceps /
instruments.
To also dissect (starting from 60/70), with strong coagulating effect, all tissues in all fields, while grazing
them with the active electrode.
SPRAY COAG
– No
‐
contact coagulation with very strong sparking. It is suitable for the following:
To coagulate (starting from 40/60), with very strong superficial effect, all tissues in all operating fields,
both by grazing the tissues with the active electrode and by touching them with surgical forceps /
instruments.
To also dissect (starting from 60/70), with strong coagulating effect, all tissues in all fields, while grazing
them with the active electrode.
PIN POINT CONTACT COAG
‐
Coagulation with medium sparking. It is suitable for the following:
For the same use of the FULG FORCED COAG current (identical starting setting), but it has a slightly
lower superficial effect.
SOFT MICRO COAG
‐
Coagulation with low sparking. It is suitable for the following:
To perform delicate coagulations (starting from 50/60) when using surgical forceps
To perform micro coagulations (starting from 10/30) when using thin active electrodes or micro needles.
It is not suitable to coagulate tissues with heavy bleeding and to cut the tissues with coagulating effect.
CURRENTS FOR BIPOLAR CUT. STANDARD BICUT
‐
Current for cut of dried or wet tissues. It is suitable for
the following:
To dissect (starting from 50/70) the tissues in Open surgery, by performing fast nips of the tissue with
forceps with thin tips.
To dissect (starting from 80/90) the tissues in Laparoscopic surgery, by using bipolar hook or bipolar
rigid needles.
BLEND BICUT
‐
Current for strongly coagulating cut (5% cut, 95% coagulation). It can be also considered a
current for coagulation only and it is very useful if users have to use a bipolar instrument for coagulation and a
bipolar instrument for vessel sealing. It is suitable for the following:
To use the traditional forceps for Open surgery (starting from 1/5 if the tips are 0, 5/1 mm, from 10/20 if
the tips are 1,5 or 2mm).
To use (starting from 30/40 according to the dimensions of the tips) forceps for laparoscopy.
To use (starting from 30/40) all scissors, for open or laparoscopic surgery.
To use (starting from 30/40) rigid needle electrodes (i.e. for turbinals or tonsillectomy)
SALINE URO
‐
GYN CUT and SALINE ARTHRO CUT currents are not suitable for this use. CURRENTS FOR
BIPOLAR COAGULATION AND SEALING OF BIG VESSELS. SOFT MICRO BICOAG
– Coagulation delicate
and precise. It is the best current for all the uses in Open / Laparoscopic surgery) and it is usable as follows:
To use the traditional forceps for Open surgery (starting from 1/5 if the tips are 0, 5/1 mm, from 10/20 if
the tips are 1,5 or 2mm).
To use (starting from 30/40 according to the dimensions of the tips) forceps for laparoscopy.
To use (starting from 30/40) all scissors, for open or laparoscopic surgery.
To use (starting from 30/40) rigid needle electrodes (i.e. for turbinals or tonsillectomy)
FORCED MICRO BICOAG
– Fast and strong coagulation. It is similar to the MICRO coagulation and it can be
used in the same way, but it produces an effect more similar to that of the monopolar coagulation. Because of this
reason it is not suitable to perform delicate coagulations by using the traditional forceps for Open surgery.