SECTION 8: Specifications and Safety
2025653-048 Revision B
Responder
™
2000
Page 96
System Overview
The Responder 2000 system has the following functions and features.
Note:
Refer to the Operator’s Manual for specific instructions and intended use.
Functions
Defibrillation (with dedicated Defibrillator operation buttons)
•
2J to 270J Biphasic waveform (can deliver 270J Biphasic shock into a 50 Ohm load)
•
Paddle connector accepts pads, paddles, or spoons
Transcutaneous cardiac pacing
40ms rectilinear current source pace pulse, 0 to 140 mA, 0 to 180 PPM, demand and asynchronous modes
ECG Monitoring through therapy electrodes, paddles, or a separate five wire ECG input
Optional Pulse Oxymetry (GE506)
Features
Color LCD to display operational, ECG, and SPO
2
information
The display is 5.67”, backlit, with a 320 x 240 resolution and over 260k colors
Trim knob for input via the display
Thermal printer for ECG, Pulse Oxymetry, Therapy Actions and Event history
•
384 pixel resolution at 25mm/s on a 60mm paper roll
•
Prints in real-time two-trace graphic, text, and raster modes
AC or rechargeable battery operation
•
AC: 90 VAC to 264 VAC, 47Hz-63Hz, 90 Watts Cont., 115 Watts intermittent
•
Battery: Lithium Ion with built in charge indicator, delivers up to 10A at 11.0V (Nominal)
Internal Defibrillator and Pacer test load
Speaker for audible indication
Serial port for software upgrade