SECTION 1: Introduction
2025653-048 Revision B
Responder
™
2000
Page 11
General Service Information
Refer to the Operator’s Manual for setup (and configuration options) required before placing the Responder 2000 into
service.
Navigation
Use the Rotary Selection knob to navigate the Responder 2000 user interface. The Rotary Selection knob is used for:
•
Scrolling through menus and sub-menus
•
Selecting soft keys
•
Setting values
Passwords
The default service password is “HEART1”. See the note under Maintenance Mode Tests in Section 5 (on page 54) for
detailed password instructions.
Battery Maintenance
The Responder 2000 has a rechargeable battery requiring periodic calibration. Calibration consists of a full charge, full
discharge, and full charge of the battery. The cycle may take up to 20 hours to complete.
To calibrate the battery using the optional external charger:
1.
Insert the battery into the external charger.
2.
Press the calibration button on the battery. The Mode light turns red to indicate calibration in progress.
To stop calibration, press the calibration button again. The Mode light turns green and the battery begins to
charge.
To calibrate the battery without the external charger:
1.
Insert the battery into the Responder 2000.
2.
Plug the Responder 2000 into an AC outlet.
3.
Allow the battery to fully charge.
4.
Disconnect the AC power cord and leave on the Responder 2000 until the battery is fully discharged (full
discharge takes about four hours).
5.
Reconnect the AC power cord and allow the battery to fully charge.