Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 71
Apex Test Load to Apex and Sternum 3.0 KVDC Test
Note:
The previous test settings were testing DC voltage. This test procedure requires setup for AC voltage.
1.
Setup the Hipot tester:
•
3.0 KVAC
•
High current limit 2.0 mA
•
Test 30 Sec.: Ramp 2.0 Sec.
•
120 VAC (±10 %) Input
2.
Plug in the Apex and Sternum Test Fixture to the ECG Paddle Connector.
3.
Connect the red hipot tester lead to the Apex and Sternum load test lead.
4.
Connect the black hipot tester lead to the Apex and Sternum lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
5.
Press
Start
(green button) on the Hipot tester. The test will take approximately two minutes.
•
If the test passes (short beep and green Pass LED is lit), continue with the next test.
•
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.
ECG to Apex and Sternum 3.0 KVDC Test
1.
Setup the Hipot tester:
•
3.0 KVAC
•
High current limit 2.0 mA
•
Test 30 Sec.: Ramp 2.0 Sec.
•
120 VAC (±10 %) Input
2.
Plug in the ECG Connector test fixture.
3.
Plug in the Apex and Sternum Test Fixture to the ECG Paddle Connector.
4.
Connect the red hipot tester lead to ECG Connector Test Fixture lead.
5.
Connect the black hipot tester lead to the Apex and Sternum lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
6.
Press
Start
(green button) on the Hipot tester. The test will take approximately two minutes.
•
If the test passes (short beep and green Pass LED is lit), continue with the next test.
•
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.
Apex and Sternum to Service Port 3.0 KVAC Test
1.
Setup the Hipot tester:
•
3.0 KVAC
•
High current limit 2.0 mA
•
Test 30 Sec.: Ramp 2.0 Sec.
•
120 VAC (±10 %) Input
2.
Plug in the Apex and Sternum Test Fixture to the ECG Paddle Connector.
3.
Plug in the Service Port Connector test fixture.
4.
Connect the red hipot tester lead to ECG Connector Test Fixture lead.
5.
Connect the black hipot tester lead to the Serial Port Connector Test lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
6.
Press
Start
(green button) on the Hipot tester. The test will take approximately two minutes.
•
If the test passes (short beep and green Pass LED is lit), continue with the next test.
•
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.