Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 81
S1: DPDT
Polarity
a) Normal
b) Reverse
S2: SPDT
Neutral
a) Closed
b) Open
S3: SPDT
Ground
a) Closed
b) Open
S4: SPDT
Test point
a) Internal
b) TP1
Position
Position
Position
Position
Maximum
Limits
(uA)
Test Description
a
b
a
b
a
b
a
b
EN/IEC
UL
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
X
X
X
X
10
10
Patient leakage current measurements: PADDLES
TP1 connected to “PADDLES” (all PADDLES connections tied together), Type CF defibrillation proof limits
Polarity: Normal
Neutral: Normal (Closed)
Ground: Normal (Closed)
X
X
X
X
10
10
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
X
X
X
X
50
50
Polarity: Normal
Neutral: Normal (Closed)
Ground: Open
X
X
X
X
50
50
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Open
X
X
X
X
50
50
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
X
X
X
X
50
50
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
X
X
X
X
10
10
Содержание Responder 2000
Страница 1: ...TM...
Страница 10: ...SECTION 1 Introduction 2025653 048 Revision B Responder 2000 Page 10 Controls and Indicators...
Страница 137: ...Contact Information 2025653 048 Revision B Responder 2000 Page 137 END OF DOCUMENT LAST PAGE...
Страница 138: ...TM...