SECTION 2: Maintenance
2025653-048 Revision B
Responder
™
2000
Page 14
Upgrade Software
Caution: Procedure Failure
Do not use this procedure to update the printer files. Updating the printer files incorrectly may cause the printer to fail. Use
the
Upgrade Printer Software
procedure on page 16.
1.
Perform the
Equipment Setup
procedure starting on page 13 until the CodeLink 2000 selection window is
displayed:
2.
Click
Choose
folder
to select all the update files in a folder or
Choose file
to select a single update file.
3.
Browse to the folder (or file) to be uploaded and click
OK
(or
Open
). While the files are updating, the status of
each file is displayed.
As the files are updated, the Shock button and then the Manual button flash on the Responder 2000. When all
files are updated, a confirmation screen is displayed
4.
Verify Pass is displayed for each file and click
OK
.
Note:
If a file fails to update, retry updating the file individually.
The Upgrade Complete Notification is displayed:
5.
Click
Yes
to upgrade more files (and repeat this procedure from step 2) or
No
to exit.