Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 74
Leads
Name
Limit
3-7
Patient Auxiliary Current DC (OS)
50 µA
3-7
Patient Auxiliary Current DC (NC)
10 µA
3-7
Patient Auxiliary Current DC OE
50 µA
3-7
Patient Auxiliary Current DC OSRM
50 µA
3-7
Patient Auxiliary Current DC NCRM
10 µA
3-7
Patient Auxiliary Current DC OERM
50 µA
1-2
Patient Leakage Current DC (OS)
50 µA
1-2
Patient Leakage Current DC NC
10 µA
1-2
Patient Leakage Current DC OE
50 µA
1-2
Patient Leakage Current DC OSRM
50 µA
1-2
Patient Leakage Current DC NCRM
10 µA
1-2
Patient Leakage Current DC OERM
50 µA
3-7
Patient Leakage Current DC (OS)
50 µA
3-7
Patient Leakage Current DC (NC)
10 µA
3-7
Patient Leakage Current DC OE
50 µA
3-7
Patient Leakage Current DC OSRM
50 µA
3-7
Patient Leakage Current DC NCRM
10 µA
3-7
Patient Leakage Current DC OERM
50 µA
8
Patient Leakage Current DC (OS)
50 µA
8
Patient Leakage Current DC (NC)
10 µA
8
Patient Leakage Current DC OE
50 µA
8
Patient Leakage Current DC OSRM
50 µA
8
Patient Leakage Current DC NCRM
10 µA
8
Patient Leakage Current DC OERM
50 µA
2.
Plug Cord from 120V, 60hz receptacle of Safety Analyzer into the wall outlet.
3.
Plug cord from
Auxiliary Power to Equipment Under Test
receptacle of Safety Analyzer into Step-up
transformer output (264VAC).
4.
Plug power cord from Responder 2000 into outlet on front of Safety Analyzer.
5.
Plug jumper cable from
ENCL
connector to
EARTH
connector.
6.
Plug the 5 ECG cable leads from ECG Cable Port into
Patient Lead
connectors
3
-
7
.
7.
Plug cable from Dfib Paddle Port into
Patient Lead
connectors
1
and
2
.
8.
(For SpO
2
equipped Responder 2000 only) Plug cable from Sp02 connector to
Patient Lead