Power Control Board and Power Supply Installation
Assembly Step
Details
Pass the three cables
through the holes in the
Lower Frame
as shown.
Secure the
Lower Frame
to
the
Front Body
with six
Torx security screws.
Figure 21: Lower Frame (Bottom View)
Route the
ECG Cable
as
shown and pass it back up
into the main compartment.
Figure 22: ECG Cable Route through Lower Frame
Install the
Fan
.
Note:
An arrow on the fan
indicates air flow direction.
Ensure the fan blows
upwards into the unit.
Figure 23: Fan Installed