Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 73
Leads
Name
Limit
8
Patient Leakage Current AC (OS)
500 µA
8
Patient Leakage Current AC (NC)
100 µA
8
Patient Leakage Current AC (OE)
500 µA
8
Patient Leakage Current AC OSRM
500 µA
8
Patient Leakage Current AC NCRM
100 µA
8
Patient Leakage Current AC OERM
500 µA
1-2
Patient Auxiliary Current AC (OS)
50 µA
1-2
Patient Auxiliary Current AC NC
10 µA
1-2
Patient Auxiliary Current AC OE
50 µA
1-2
Patient Auxiliary Current AC OSRM
50 µA
1-2
Patient Auxiliary Current AC NCRM
10 µA
1-2
Patient Auxiliary Current AC OERM
50 µA
3-7
Patient Auxiliary Current AC (OS)
50 µA
3-7
Patient Auxiliary Current AC (NC)
10 µA
3-7
Patient Auxiliary Current AC (OE)
50 µA
3-7
Patient Auxiliary Current AC OSRM
50 µA
3-7
Patient Auxiliary Current AC NCRM
10 µA
3-7
Patient Auxiliary Current AC OERM
50 µA
1-2
Mains on Applied Part (SFC)
50 µA
1-2
Mains on Applied Part (SFCRM)
50 µA
3-7
Mains on Applied Part SFC
50 µA
3-7
Mains on Applied Part SFCRM
50 µA
8
Mains on Applied Part SFC
5000 µA
8
Mains on Applied Part SFCRM
5000 µA
1-2
Patient Leakage Current DC (OS)
50 µA
1-2
Patient Leakage Current DC (NC)
10 µA
1-2
Patient Leakage Current DC (OE)
50 µA
1-2
Patient Auxiliary Current DC OSRM
50 µA
1-2
Patient Auxiliary Current DC NCRM
10 µA
1-2
Patient Auxiliary Current DC OERM
50 µA