Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 53
Maintenance Menu Reference
Caution: Loss of Patient Data
Do not enter Maintenance mode while the Responder 2000 is connected to a patient. Patient data may be lost. Perform
any Maintenance mode procedures outside the patient vicinity.
Note:
The Responder 2000 battery must have at least a 20% charge before performing these procedures.
The following Maintenance Menu items are listed as a reference for quick troubleshooting or verification and are not
intended as a comprehensive operational or safety checkout. The entire checkout procedures listed above must be
performed before returning the Responder 2000 to the customer.
Note:
A password is required to access the Maintenance menu. The default password is HEART1, but may have been
changed by the facility. To enter a password:
1.
Select
Maintenance
from the
System
menu. The password field is displayed, highlighted.
2.
Click to highlight the first character field.
3.
Click the highlighted field, and then rotate the encoder button to the first letter of the password.
4.
Click to accept the letter and then rotate to the next field.
5.
After all password characters are entered, rotate one space past the last character field. The border around the
password field highlights.
6.
Click to highlight the password field and then rotate and click
Accept
.