Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 60
Visual Inspection
Visually inspect the unit and accessories for foreign substances, damage, cracks, bent or discolored pins, broken parts, or
extreme wear, including:
•
Cracked, bulging, or otherwise damaged case
•
Cracked, loose, or bent connectors or plugs
•
Rattling when gently shaken
•
Discoloration or signs of overheating
•
Obvious scratches or damage to the display
•
Loose battery or paddle clips
•
Loose printer door
Operational Tests
Initial Power Up
Perform this procedure for the following scenarios:
•
With a charged battery (at least 25%) installed and connected to AC power
•
With the battery removed and connected to AC power
•
With a charged battery (at least 25%) and AC power disconnected
Turn on the Responder 2000 and verify the following during power up:
•
No alarms (at any time during power up or operation)
•
The speaker emits a short beep, then is silent
•
The following indicators momentarily light:
o
Service Required: Red
o
Charge: off
o
Manual: Blue
o
Shock: Red
•
The Responder 2000 is ready for use in about 4 seconds.
The AC Power LED and Battery Charging LED are lit when a battery is installed and the Responder 2000 is connected to
AC power
•
The AC Power LED is not lit and Battery Charging LED is not lit when a battery is installed and the Responder
2000 is not connected to AC power
Содержание Responder 2000
Страница 1: ...TM...
Страница 10: ...SECTION 1 Introduction 2025653 048 Revision B Responder 2000 Page 10 Controls and Indicators...
Страница 137: ...Contact Information 2025653 048 Revision B Responder 2000 Page 137 END OF DOCUMENT LAST PAGE...
Страница 138: ...TM...