Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 75
Test Procedure
Note:
This procedure assumes the use of a Metron QA-90 Safety Analyzer. If an equivalent safety analyzer is used, refer
to the Operator’s manual for detailed instructions.
1.
Press Responder 2000
Power
button to turn on.
2.
Turn on the Safety Analyzer and wait for the self test to complete.
3.
If previously saved, select the stored Responder 2000 parameters.
4.
Press
START
on the Safety Analyzer to begin the tests.
5.
After test sequence is complete, verify TEST PASSED is displayed on the Safety Analyzer.
Alternate Safety Test Procedures
The following procedures are provided in case the recommended test procedures (above) cannot be performed.
Test Equipment
Number
Equipment Description
Qty
Fluke Model Index 2 Sp02 simulator
1
Fluke Model ECG Simulator Defib analyzer or adequate
1
Metron QA-90 Safety Analyzer or adequate
1
MarqIII-Kit
Marq-3 EKG Simulator or adequate
1
2025269-003
RESPONDER 2000/PH ECD PACER TESTER
1