SECTION 8: Specifications and Safety
2025653-048 Revision B
Responder
™
2000
Page 101
Detection Lead:
Through either 3-lead and 5-lead patient cable, pads, or
paddles
Lead Selection:
3-Lead Cable: (I, II, III) and 5-Lead (I, II, III, aVL, aVR,
aVF, and V), PADS: (Modified Lead II)
Heart Rate Range:
25 to 300 bpm
Display Frequency Response (ECG Leads):
0.5 to 100 Hz (± 10 %)
Display Frequency Response (paddles):
3 Hz to 33 Hz (± 10 %); (-3db)
ECG Effective Sampling Rate:
1000 samples per second (ECG Leads), 128 samples per
second (paddles)
Post Defibrillation Recovery:
8 seconds
Heart Rate Alarms:
Off, 5-300 bpm
Leads Off Sensing
DC current; Sensing Leads <0.1 uA; Reference Lead
<1uA; when all leads are connected
ECG Leads and paddles leakage current
Less than 10uA in normal; Less than 50uA in single fault
condition
Dynamic Range: Input ECG signal amplitude:
± 5mV
Dynamic Range: DC Offset voltage:
± 500 mV ECG from ECG cable; ± 1000 mV (ECG from
pads, paddles and spoons
Asystole Threshold
0.2mV (±0.1mV)
Rechargeable Battery
Battery Voltage:
11.1 V Nominal
Chemistry:
Lithium-ion
Compatibility:
Compatible with Responder 2000
Battery Capacity:
50 shocks, or 240 minutes monitoring time, or 72 minutes
monitoring time with pacing
Battery Charge Time:
8 hours in Responder 2000, 4 hours in external charger,
20 hours for calibration cycle in external charger
Battery Standby:
6 months
Battery Life:
2.5 years or 300 Battery charge-discharge cycles,
whichever comes first
Battery Weight:
1 lb. 3 oz; .54 kg
AC Power Supply
Universal input:
100V to 240V~
50Hz-60Hz
200VA
Printer
Speed
25 mm/S feed rate
Paper Size
60 mm paper width