background image

GE Healthcare

(FEBRUARY 28, 2019)
COVER.FM

P R E L I M I N A R Y

LOGIQ S8/LOGIQ E8

Service Manual

Document Number: 

5394227

Revision: 

12

Содержание LOGIQ S8

Страница 1: ...GE Healthcare FEBRUARY 28 2019 COVER FM P R E L I M I N A R Y LOGIQ S8 LOGIQ E8 Service Manual Document Number 5394227 Revision 12...

Страница 2: ...al is a reference for the LOGIQ S8 LOGIQ E8 and LOGIQ S8 Vet ultrasound systems Hereafter listed as LOGIQ S8 All information provided in this manual is relevant for all three systems unless otherwise...

Страница 3: ...QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESPECT DE CET AVERTISSE...

Страница 4: ...TE DEVIDO A CHOQUES EL TRICOS MEC NICOS OU OUTROS ESTE MANUAL DE ASSIST NCIA EST DISPON VEL APENAS EM INGL S SE QUALQUER OUTRO SERVI O DE ASSIST NCIA T CNICA QUE N O A GEHC SOLICITAR ESTES MANUAIS NOU...

Страница 5: ...TERIAALIA TARVITTAVAN K NN KSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMM RT NYT T M N HUOLTO OHJEEN MIK LI T T VAROITUSTA EI NOUDATETA SEUR...

Страница 6: ...HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICK HO PO KOZEN NEBO JIN MU RIZIKU DENNE SERVICEMANUAL FINDES KUN P ENGELSK HVIS EN KUNDES TEKNIKER HAR BRUG FOR...

Страница 7: ...SU EISTAS D L ELEKTROS SM GIO MECHANINI AR KIT PAVOJ DENNE SERVICEH NDBOKEN FINNES BARE P ENGELSK HVIS KUNDENS SERVICELEVERAND R TRENGER ET ANNET SPR K ER DET KUNDENS ANSVAR S RGE FOR OVERSETTELSE IK...

Страница 8: ...AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR OVAJ PRIRU NIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLESKOM JEZIKU A...

Страница 9: ...PROUZRO ENE STRUJNIM UDAROM MEHANI KIM I DRUGIM OPASNOSTIMA T TO SERVISN PR RU KA JE K DISPOZ CII LEN V ANGLI TINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK AKO ANGLI TINU POSKYTNUTIE PREKLA...

Страница 10: ...N NG L ZCE DI INDAK B R D LDE OLMASINI STERSE KILAVUZU TERC ME ETT RMEK M TER N N SORUMLULU UNDADIR SERV S KILAVUZUNU OKUYUP ANLAMADAN EK PMANLARA M DAHALE ETMEY N Z BU UYARININ G Z ARDI ED LMES ELEKT...

Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...

Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...

Страница 13: ...ical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel...

Страница 14: ...basic service software does not extend to or cover this software or related documentation If you are a GE Healthcare employee or a customer who has entered into such a license agreement with GE Health...

Страница 15: ...ptember 2015 Update 8 November 2015 Update BW Printer setup 9 Februry 2016 Add UDI label information 10 Februry 2017 Update adding R4 information 11 June 2017 Revise with Service Note information 12 O...

Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...

Страница 17: ...tions used in this manual 1 7 Standard Hazard Icons 1 8 Product Icons 1 10 Safety Considerations 1 15 Contents in this section 1 15 Introduction 1 15 Human Safety 1 15 Mechanical Safety 1 17 Electrica...

Страница 18: ...RE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 2 Electrostatic Discharge ESD Prevention 1 30 Customer Assistance 1 31 Contact Information 1 31 System Manufacturer 1 32 Authorized EU Represen...

Страница 19: ...Requirements 2 3 EMI Limitations 2 4 Probe Environmental Requirements 2 6 Time and Manpower Requirements 2 6 Facility needs 2 7 Purchaser Responsibilities 2 7 Required Facility Needs 2 8 Privacy and...

Страница 20: ...S8 3 9 Packing materials recycling information 3 10 Preparing for setup 3 11 Verify Customer Order 3 11 Physical inspection 3 11 EMI Protection 3 11 Completing the Set Up 3 12 System Specifications 3...

Страница 21: ...the printer to print reports 3 65 PC printer setup 3 65 Connectivity Setup 3 66 Connectivity Introduction 3 66 The Dataflow Concept 3 66 Stand alone LOGIQ S8 3 68 LOGIQ S8 PC within a Sneaker Net 3 6...

Страница 22: ...ing Request for Service RFS 3 90 Verifying InSite Connection CSD 3 91 Verifying InSite Connection VCO 3 92 Agent Configuration R4 2 5x and later 3 93 Overview 3 93 Verifying back office connection CSD...

Страница 23: ...rvice Platform Confirmation R1 to R4 4 17 Functional Checks 4 18 Overview 4 18 Contents in this section 4 18 Preparation 4 18 Basic Controls 4 19 Performance Tests 4 22 B Mode Checks 4 23 M Mode Check...

Страница 24: ...CE MANUAL 8 Volume Navigation Option 4 46 FibroScan module Option 4 47 Application Turnover Check List 4 48 Software Configuration Checks 4 48 Power supply test and adjustments 4 48 Power Supply Test...

Страница 25: ...tion 5 11 Power Loss description 5 12 Software Overview 5 13 Purpose of this section 5 13 Contents in this section 5 13 LOGIQ S8 software 5 13 Software Options 5 14 Hardware Options 5 17 LOGIQ S8 R4 F...

Страница 26: ...ts Overview 6 1 Purpose of this chapter 6 1 Monitor Adjustment 6 2 19 LCD Monitor 6 2 22 OLED and 23 Wide Monitor Adjustment 6 8 Volume Navigation Calibration Procedure 6 10 Purpose of this section 6...

Страница 27: ...ion 7 5 Collecting Vital System Information 7 6 Collecting a Screen Capture with Logs 7 9 Capturing Service Logs with ALT D 7 12 Screen captures 7 13 Capturing Network Logs with Network Sniffer 7 14 S...

Страница 28: ...7 74 Universal Video Converter Troubleshooting 7 75 4D Option Trouble shooting 7 75 Power Assistant Option Trouble Shooting 7 75 Extended Battery Option Trouble Shooting 7 76 Mode encoder works twice...

Страница 29: ...HCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 13 CHAPTER 8 Renewal Parts Overview 8 1 Contents in this chapter 8 1 Optional Peripherals and Accessories 8 2 Printers 8 2 Probes 8 4 Power...

Страница 30: ...unctional checks 9 8 Cleaning 9 10 Physical inspection 9 13 Probe maintenance 9 14 Using a Phantom 9 16 Electrical Safety Tests 9 17 Safety test overview 9 17 GEHC Leakage Current Limits 9 18 Outlet T...

Страница 31: ...227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 15 When There s Too Much Leakage Current 9 35 Chassis Fails 9 35 Probe Fails 9 35 Peripheral Fails 9 35 Still Fails 9 35 ECG Fails 9 36 Ultrasound Equipment Qua...

Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...

Страница 33: ...eld Service trainers for the agreed upon time period This service manual provides installation and service information for the LOGIQ S8 ultrasound system as shown in Table 1 1 Contents in this service...

Страница 34: ...ure checklist 4 Functional Checks Contains functional checks that must be performed as part of the installation or as required during servicing and periodic maintenance 5 Components and Functions Theo...

Страница 35: ...SERVICE MANUAL Section 1 2 Service manual overview 1 3 1 2 2 Typical users of the Service Manual GE Service Personnel Setup maintenance etc Hospital Service Personnel Architectural Planners Installati...

Страница 36: ...configuration Y U U N N 5418104 LOGIQ S8 Console USA configuration Y U U N N 5418105 LOGIQ S8 Console CANADA configuration Y U U N N 5478057 LOGIQ S8 R2 110V N Y U N N 5478058 LOGIQ S8 R2 220V N Y U...

Страница 37: ...e Configurations and Hardware Software Compatibility Upgrade Options Y Original U Upgrade available N Not supported U Upgrade available No LCD upgrade Table 1 4 LOGIQ E8 Model ship only for People s R...

Страница 38: ...Doppler M Mode Color M Mode PW CW and 3D 4D Tissue Velocity imaging Volume Guided Ultrasound Elastography and Contrast applications The fully digital architecture of the LOGIQ S8 LOGIQ E8 allows optim...

Страница 39: ...rd is present that can cause property damage but has absolutely no personal injury risk a NOTICE is used Notice Example Disk drive may crash NOTE Notes are used to provide important information about...

Страница 40: ...or Danger Warning or Caution as seen throughout this chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause h...

Страница 41: ...tandard Icons that indicate that a special procedure is to be used Other hazard icons make you aware of specific procedures that should be followed Table 1 6 Standard Icons Indicating a Special Proced...

Страница 42: ...es only Prescription Requirement label Rear of the system CE Mark The CE Mark of Conformity indicates this equipment conforms with the Council Directive 93 42 EEC Rear of the system Authorized Europea...

Страница 43: ...icates the power off position of the mains power breaker Rear of the system Mains ON indicates the power on position of the mains power breaker Rear of the system ON indicates the power on position of...

Страница 44: ...or disposed of in accordance with local or country laws The letters below the separate collection symbol indicate whether certain elements Pb Lead Cd Cadmium Hg Mercury are contained in the battery To...

Страница 45: ...rm outside the published specifications Keep power to these devices turned off when near this equipment Rear of the system This product consists of devices that may contain mercury which must be recyc...

Страница 46: ...Mark of Products Circulation in the market of the Customs Union member states Under the rating plate Every system has a unique marking for identification the Unique Device Identification UDI Label The...

Страница 47: ...ly personnel who have participated in a LOGIQ S8 Training are authorized to service the equipment DANGER DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION W...

Страница 48: ...HERE ARE NO TEST POINTS TO VERIFY ISOLATION THE BLUE LIGHT ON THE POWER BUTTON WILL TURN OFF WARNING WARNING RISK OF ELECTRICAL SHOCK SYSTEM MUST BE TURNED OFF AND DISCONNECTED FROM POWER SOURCE CORD...

Страница 49: ...ready for use Care must be used when moving it or replacing its parts Failure to follow the precautions listed could result in injury uncontrolled motion and costly damage ALWAYS Use the handle to mo...

Страница 50: ...unit in a vehicle see 4 3 8 3 Transporting the System on page 4 16 CAUTION Use Protective Glasses during drilling filing and during all other work where eyes need protection CAUTION Use Safety Shoes...

Страница 51: ...is correct handling of probes is important and the following points should be noted Do not drop a probe or strike it against a hard surface as this may damage the transducer elements acoustic lens or...

Страница 52: ...a potential difference between the protective earth and any line of the computer network including the shield In this case the only way to operate the system safely is to use an isolated signal link...

Страница 53: ...pecial main outlet provided for the electrical safety of the system CAUTION Auxiliary equipment with direct main connection requires galvanic separation of the signal and or control leads NOTICE The s...

Страница 54: ...p on it DO NOT use a damaged battery DO NOT apply solder to a battery DO NOT connect the battery to an electrical power outlet DO NOT immerse the battery in water or allow it to get wet DO NOT place t...

Страница 55: ...e labels on the cover The LOGIQ S8 comes equipped with product labels and icons These labels and icons represent pertinent information regarding the operation of the unit NOTE The rating label is loca...

Страница 56: ...Labels Locations Figure 1 2 Label location R4 and later system NOTE Depending on country Gender Label is attached 1 LCD Caution label 2 LOGIQ S8 rating label 3 Gender Caution only for India China Kore...

Страница 57: ...94227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 1 5 Labels Locations 1 25 1 5 1 1 WLAN label location if WLAN option installed Figure 1 3 WLAN Label location R1 to R3 system Figure 1 4 WLAN Label lo...

Страница 58: ...on the rating plate Figure 1 5 Rating plate and UL label example NOTE The rating label is located on the rear of R1 R2 R3 system Figure 1 6 Rating plate and UL label R4 example 1 Catalogue number 2 M...

Страница 59: ...N HANDLING TESTING AND ADJUSTING WARNING WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ S8 LOGIQ E8 SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED EVEN WH...

Страница 60: ...isolation All potentially hazardous stored or residual energy is relieved NOTICE Energy Control and Power Lockout for LOGIQ S8 When servicing parts of the system where there is exposure to voltage gre...

Страница 61: ...that items what were saturated and or dripping with human blood that are now caked with dried blood or which were used or intended for use in patient care are regulated medical waste for transportatio...

Страница 62: ...c discharge radiated and conducted RF fields magnetic fields and power line transient requirements as mentioned in IEC 60601 1 2 NOTE For CE Compliance it is critical that all covers screws shielding...

Страница 63: ...ndary image capture cine loop NOTE Restart the application before resuming clinical scanning Table 1 8 Phone Numbers for Customer Assistance Location Phone Number USA GE Healthcare Ultrasound Service...

Страница 64: ...r GE Ultrasound Korea 9 Sunhwan ro 214beon gil Jungwon gu Seongnam Si Gyeonggi do Korea 82 0 31 740 6436 Table 1 10 Authorized EU Representative Manufacturer TEL FAX Number The location of the CE mark...

Страница 65: ...iew 2 1 1 Purpose of this chapter This chapter provides the information required to plan and prepare for the setup of a LOGIQ S8 Included are descriptions of the facility and electrical needs to be me...

Страница 66: ...system installation updates and repairs However operator and patient comfort may be optimized if the room light is subdued and indirect Therefore a combination lighting system dim bright is recommend...

Страница 67: ...eripherals Power Consumption nominal 900 VA including all on board peripherals 2 2 2 3 Inrush Current Inrush current is not a factor to consider due to the inrush current limiting properties of the po...

Страница 68: ...that interference will not occur in a particular installation Possible EMI sources should be identified before the unit is installed Electrical and electronic equipment may produce EMI unintentionall...

Страница 69: ...ow radio frequencies to interfere with the ultrasound signals Replace broken RF gaskets If more than 20 or a pair of the fingers on an RF gasket are broken replace the gasket Do not turn ON the unit u...

Страница 70: ...soon as possible If possible allow six weeks before delivery for enough time to make necessary changes Table 2 4 Probe Environmental Requirements Probe Operation Storage Transport Temperature 2D probe...

Страница 71: ...erformed by qualified personnel The products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrical work on these...

Страница 72: ...roposed location for the LOGIQ S8 is at least 0 2 m 0 67 ft from the walls to enable cooling Power outlet and place for any external peripheral are within 2 m 6 5 ft of each other with peripheral with...

Страница 73: ...t Sink with hot and cold running water Receptacle for bio hazardous waste for example used probe sheaths Emergency oxygen supply Storage area for linens and equipment Nearby waiting room lavatory and...

Страница 74: ...e the transfer of images to workstations for viewing or transferring images to remote printers As an added benefit transferring images in this manner frees up the on board LCD and peripherals enabling...

Страница 75: ...Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept LOGIQ S8 SN Type Floor Room REV Comments CONTACT INFORM...

Страница 76: ...providing the device with extra protective earth will be required in order to meet UL60601 1 and IEC60601 1 IEC60601 1 1 standards for electrical leakage 2 4 1 Patient Vicinity UL60601 1 USA 2 12 20D...

Страница 77: ...1 2005 Such an area is an environment in which medical diagnosis monitoring or treatment is carried out It is very difficult to attach unique dimensions to the PATIENT ENVIROMENT In practice a distanc...

Страница 78: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 2 14 Section 2 4 Environmental Dangers This page was intentionally left blank...

Страница 79: ...he unit probes and external peripherals for electrical safety are included in this procedure 3 1 2 Contents in this chapter 3 1 Overview 3 1 3 2 Setup reminders 3 2 3 3 Receiving and unpacking the equ...

Страница 80: ...site has been prepared the average installation time required is shown in table below Table 3 1 Average Installation Time Description Average Setup Time Comments Unpacking the scanner 0 5 hours Setup...

Страница 81: ...PANEL EXTENDED MOVE THE OPERATOR CONTROL PANEL TO IT S CENTERED AND LOCKED POSITION REFER TO Section 8 2 3 Returning Shipping Probes and Repair Parts on page 8 4 CAUTION CAUTION TO PREVENT ELECTRICAL...

Страница 82: ...after transportation of the system with the monitor in the folded down position it is highly recommended to place the monitor in the upright position and to leave it in this position for a period of l...

Страница 83: ...TEM WITH THE FRONT HANDLE OF THE USER INTERFACE OPERATOR PANEL CAUTION Two people are needed to unpack the unit because of its weight Attempts to move the unit considerable distances or on an incline...

Страница 84: ...the tilt indicator Is the tilt Indicator red colored 1 inside the middle of the indicator If YES The tilt Indicator has been activated Continue with the instructions in Section 3 3 2 3 Damage in trans...

Страница 85: ...is undamaged To make it easier to detect if the handling during transportation has been improper a set of Tilt Shock indicators have been attached to the transportation box 3 3 2 4 2 Position of the T...

Страница 86: ...PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 8 Section 3 3 Receiving and unpacking the equipment 3 3 3 Transportation Box Label Figure 3 2 Label 1 Figure 3 3 Label 2 Figur...

Страница 87: ...ng the LOGIQ S8 ultrasound scanning unit Images are ONLY for reference wear proper PPE when handling packaging gloves safety shoes etc Table 3 5 Unpacking Procedure Step Task 1 Cut the straps around t...

Страница 88: ...rd The inner reinforcements are made of Ethafoam Polyethylene foam The plastic foil is made of LDPE Low Density Polyethylene 7 Unlock the casters and align the orientation of the 4 caster s direction...

Страница 89: ...fer to 3 3 2 3 Damage in transportation on page 3 6 in the beginning of this manual 3 5 3 EMI Protection This unit has been designed to minimize the effects of Electo Magnetic Interference EMI Many of...

Страница 90: ...n max 1150 mm 45 3 in min 500 mm 19 7 in Keyboard 620 mm 24 4 in Caster 865 mm 34 1 in max 795 mm 31 3 in Caster 85 kg 187 4 lbs Table 3 7 LOGIQ S8 R4 and later without Peripherals Height Width Depth...

Страница 91: ...a label on the back of the system on lower rear frame of the LOGIQ S8 3 6 2 2 Electrical Specifications In the table below the electrical specifications for LOGIQ S8 include LCD and on board periphera...

Страница 92: ...e signal input part or signal output part of the LOGIQ S8 system configures a medical system and is therefore responsible to ensure that the system complies with the requirements of the valid version...

Страница 93: ...connector 2 Put the probe in the probe holder Use the integrated cable management hook to wrap the cord 3 Ensure that the probe locking handle is at horizontal position 4 Hold the probe connector ver...

Страница 94: ...LOGIQ S8 system Figure 3 8 DLP RS Connector 1 Hold the probe connector and TEE DLP RS adapter with the cable pointing DLP connector lock knob 2 Turn the RS connector lock knob to release side Figure 3...

Страница 95: ...d battery charging the LED is blinking and turn off when the battery is fully charged DANGER ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE PROTECTIVE GROUNDING CONNECTOR DANGER NEVER U...

Страница 96: ...ircuit Breaker is in OFF position if not switch it OFF Figure 3 10 Circuit breaker and main power cable at rear of system 8 Connect the Power Cable s other end male plug to a hospital grade mains powe...

Страница 97: ...the scanner During a normal boot you may observe Power is distributed to Peripherals Operator Panel control panel Monitor Front End and Back End Processor The Back End Processor and rest of the scanne...

Страница 98: ...are acknowledging that you are authorized to do so and are trained in appropriate use and regulatory guidelines Figure 3 12 Login Window Example of R4 2 5x After you have established your password fol...

Страница 99: ...down The power button turns on blue indicated Standby mode refer to Section Figure 3 11 Power Button on page 3 19 For the system with R4 2 5x and later press Power On Off button three 3 times to shut...

Страница 100: ...s before turning it on again The system may not be able to boot if power is recycled too quickly NOTICE The The system will go to Power Assistant Mode Optional R2 or later if above procedures are skip...

Страница 101: ...into the sleep mode enter the scan screen and lightly press the Power On Off switch once The System Exit window is displayed 2 Select Sleep The system performs sequence to go to SLEEP mode 3 Unplug th...

Страница 102: ...first logging onto the system 1 Type your name in the Operator field 2 Press the Change Password button The Change Password pop up displays Figure 3 13 Change Password Pop Up 3 Type the following Pas...

Страница 103: ...policies To create password policies 1 Navigate to Utility Admin Logon 2 Select Enable Password Policies The system prompts the user for a password at logon Figure 3 14 Policies Window Example of R4...

Страница 104: ...3 Enter password using the defined policies 4 If needed select User Must Change Password The user will be prompted to change the password on the next logon 5 Under Group Membership select the groups...

Страница 105: ...Group List select from the list of groups with pre determined access is available to select for each new user b To view access rights for these groups highlight a group and then under Group Rights vie...

Страница 106: ...in minutes The screen will appear completely black when locked timeout 4 Click Save 5 Press or move any button trackball on the operator panel to display Login window 6 On the Login window enter the...

Страница 107: ...Logon after starting system This is not the user password this is system access for all users 5 Recovery key is created this can be saved to a USB flash drive 6 Insert a USB flash drive in the USB por...

Страница 108: ...overy key The generated key will always be different 13 Select Ok to continue Encryption of data begins This may take up to 90 minutes the first time After the initial encryption there is no long proc...

Страница 109: ...ck Secure the password and recovery key including USB memory It is recommended the recovery key is stored in a USB flash drive and printed out If both password and recovery key are lost you will not b...

Страница 110: ...control adjust connectivity settings starting with Section 3 11 Connectivity Setup on page 3 66 3 8 2 System Configuration For complete instructions refer to the Basic User Manual Chapter 16 R1 to R3...

Страница 111: ...u 3 Set Hospital name Department name Language Units Time Date format as needed 4 To save the changes select Save on the bottom left Select Exit to return to scanning In some cases you may need to reb...

Страница 112: ...ameter Description Time Format Select the time format 12 Hr AM PM or 24 Hr Date Format Select the date format dd mm yyyy mm dd yyyy or yyyy mm dd Default Century Select the default century for the sys...

Страница 113: ...ord is specific for each LOGIQ S8 3 8 3 1 Installing a Software Option 1 Before you install a software option you must login 2 From the Touch Panel select Utility Admin System Admin 3 Enter the new op...

Страница 114: ...R3 and later the Option String can be imported from the media 5 Press Save to save the new setting 6 Restart to save and activate the settings and adjustments you have done so far Figure 3 25 Option...

Страница 115: ...please review Section 1 4 6 Auxiliary Devices Safety on page 1 20 NOTE Normally auxiliary devices and peripherals come pre installed with the system WARNING WARNING WARNING After each installation the...

Страница 116: ...CAUTION CAUTION When you connect the Off board On board UVC to the Secondary Patient Monitor If the secondary patient monitor to be located inside of the patient environment to be connected to the app...

Страница 117: ...CD Monitor cont d 3 9 1 1 WIthout UVC Figure 3 26 Connection Scheme Secondary LCD Monitor without UVC HDMI to DVI HDMI to HDMI cable connected to the rear side of the 2nd Monitor Secondary Monitor App...

Страница 118: ...nitor with Off board UVC Secondary Monitor Appropriate Transformer A Secondary Patient Monitor MUST NEVER be connected to the LOGIQ S8 s mains supply directly CAUTION CAUTION Output connection Followi...

Страница 119: ...Monitor without UVC Secondary Monitor Appropriate Transformer A Secondary Patient Monitor MUST NEVER be connected to the LOGIQ S8 s mains supply directly CAUTION CAUTION Output connection Following O...

Страница 120: ...tional Peripherals Peripheral Connection 3 9 2 Connecting the Off Board On Board UVC 3 9 2 1 Connecting the Off Board UVC 1 Power OFF Shutdown the system as described in Section 3 7 Power On Off on pa...

Страница 121: ...tional Peripherals Peripheral Connection 3 43 3 9 2 1 Connecting the Off Board UVC cont d Figure 3 30 Off Board UVC Connection for R3 and later 3 When all the cables are connected press the Power ON s...

Страница 122: ...9 2 2 Connecting the On Board UVC 1 Power OFF Shutdown the system as described in Section 3 7 Power On Off on page 3 17 2 Connect the UVC according to correct connection scheme Figure 3 31 R2 On Boar...

Страница 123: ...Power ON Boot up the system as described in Section 3 7 Power On Off on page 3 17 3 9 2 3 UVC functional check Procedure 1 Connect output DVI or D sub or S video Composite on UVC to the external monit...

Страница 124: ...itch should be directly connected to any accessible USB port on the system e g on rear of the system NOTE Connection of the Footswitch is always the same no differences between PC Motherboard version...

Страница 125: ...onal Peripherals Peripheral Connection 3 47 3 9 4 ECG connector Connect the ECG cable to the ECG connector on the monitor arm or on the side of the operator panel depending on the system version Figur...

Страница 126: ...emarks and Hints when using external USB3 0 Devices R4 or later NOTICE NOTICE Before connecting an USB device please read Section 3 9 5 1 General Remarks and Hints when using external USB Devices on p...

Страница 127: ...nnector 3 When all the cables are connected press the power button to ON on the printer 4 Power ON Boot up the system as described in Section 3 7 Power On Off on page 3 17 All software drivers are pre...

Страница 128: ...E PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 50 Section 3 10 Printer Installation 3 10 1 1 Connection Scheme Black White Printer Figure 3 37 R1 and R2 Black White Printe...

Страница 129: ...IETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 3 10 Printer Installation 3 51 3 10 1 1 Connection Scheme Black White Printer cont d Figure 3 38 Black White Printer connecti...

Страница 130: ...below for printer setting Table 3 3 Settings for UP D897 D898 Recommended Setting Paper size 1920x1280 Windows Orientation Portrait Utility Orientation Portrait Avoid using WIDE setting In Utility Sys...

Страница 131: ...ter Properties Click Properties 4 Select Properties from Printer pull down menu 5 Click Printing Preferences at the bottom of Properties Window 6 Select Layout tab and select Paper 1920x1280 Orientati...

Страница 132: ...ERVICE MANUAL 3 54 Section 3 10 Printer Installation 3 10 1 3 BW Printer shape differences Setting 1920x1280 Portrait in Printer Properties Portrait in Utility Connectivity Service Table 3 4 Print exa...

Страница 133: ...MANUAL Section 3 10 Printer Installation 3 55 3 10 1 3 BW Printer shape differences cont d Setting 1920x1280 Landscape in Printer Properties Landscape in Utility Connectivity Service Table 3 5 Print...

Страница 134: ...ION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 56 Section 3 10 Printer Installation 3 10 1 3 BW Printer shape differences cont d Setting 960x1280 Print output differ by BW Printer Model to orientat...

Страница 135: ...Medium Previous Service Manual recommendation 1920x1280 P P Medium Recommended 1920x1280 L L Large UP D898 960x1280 P P Very small Unfit for use 1920x1280 P P Medium Recommended 1920x1280 L L Large R3...

Страница 136: ...the cables are connected press the power button to ON on the Color printer 4 Power ON Boot up the system as described in Section 3 7 Power On Off on page 3 17 All software drivers are pre installed f...

Страница 137: ...LTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 3 10 Printer Installation 3 59 3 10 2 1 Connection Scheme Color Printer Figure 3 40 R1 and R2 Color Printer conn...

Страница 138: ...PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 60 Section 3 10 Printer Installation 3 10 2 1 Connection Scheme Color Printer cont d Figure 3 41 Color Printer connection for...

Страница 139: ...m Printer pull down menu 3 Click Printing Preferences at the bottom of Properties Window 4 Select the Paper tab and select Recommended Paper UPC 24LA large UPC 24SA small NOTE Paper UPC 21L and UPC 21...

Страница 140: ...ings cont d 5 Select the Graphics tab From the Color Adjust pop up menu select a Color Balance Cyan 0 Magenta 0 Yellow 0 b Gamma Select Gamma 1 Figure 3 43 Graphics page Color Balance Gamma Select c C...

Страница 141: ...nter Installation 3 63 3 10 2 2 UP D25MD Printer Settings cont d 6 For saving the adjusted printer settings click Apply and then OK 7 Finally close the Printers window with the close button and exit S...

Страница 142: ...from the Printer pull down Properties menu For the UP D897 D898 printer select Portrait as orientation 5 Type the printer name in the Name field This name is used on the Button screen After you selec...

Страница 143: ...Utility System Peripherals 2 Select the printer from Default Printer pull down menu Figure 3 45 Report Printer Setup 3 Press Save 4 Press Print on the Report screen to print the report 3 10 5 PC prin...

Страница 144: ...S8 PC within a Sneaker Net on page 3 68 LOGIQ S8 and DICOM server in a network see Section 3 11 5 Connection between LOGIQ S8 and DICOM Server on page 3 68 3 11 2 The Dataflow Concept Communication be...

Страница 145: ...Images are stored to internal hard drive The local database is used for patient archiving Afterwards images are stored to a DVD CD or external USB device etc Search in the DICOM Modality Worklist the...

Страница 146: ...connection needed 3 11 5 Connection between LOGIQ S8 and DICOM Server In this configuration the LOGIQ S8 is configured to work with a DICOM server in a network environment Usually this will be the ho...

Страница 147: ...connector on the LOGIQ S8 2 Connect the other end to the network connector to the Workstation see the Workstation Service Manual 3 11 6 2 Connection via a Peer to Peer Network You will need a network...

Страница 148: ...tion Wired network connected Wired network disconnected Wireless network connected Wireless network disconnected Wireless network limited connected No IP address obtained Empty IP means Acquiring Netw...

Страница 149: ...2 R4 2 5x and later Table 3 8 Network icon description Ethernet Active Ethernet Error Ethernet Active Spooler Active Ethernet Active Spooler Error Spooler Active Spooler Active Error Spooler Inactive...

Страница 150: ...d by individual fields This greatly improves query capability DICOM Structured Reporting creates coded clinical data that can be used for clinical research outcomes analysis and disease management DIC...

Страница 151: ...a dataflow customizes the ultrasound system to work according to the services associated with the selected dataflow 5 Button allows you to assign a pre configured output service or a set of output se...

Страница 152: ...ollowing DO NOT enable DHCP Type the IP Address acquire unique static IP address from hospital network administrator Subnet Mask and Default Gateway if applicable NOTE Do not set up the system with DH...

Страница 153: ...P Address Subnet Mask and Default Gateway when the user chooses to use DHCP IP Address Type the IP Address of the Ultrasound system NOTE IP stands for Internet Protocol Every device on the network has...

Страница 154: ...miley face frowns then the connection has not been made Check the device name and IP address Table 3 11 Device Preset Parameter Description Add Remove Press Add to add a new device press Remove to del...

Страница 155: ...e lists information about destination devices You can select from a list of currently existing devices 2 Service Type to Add lists information about services for the destination device You can add ser...

Страница 156: ...etrieve DICOM Storage Commitment DICOM Worklist Standard Print Video Capture Save As Network storage USB Quick Save R2 and later Email to MMS R4 and later Table 3 12 Service Parameters Common Service...

Страница 157: ...vided by IT Select Utility Connectivity on the touch screen then select the Device Tab On the Device Tab click Add enter New Device Name and IP Address Click Save Select Ping to verify the icon change...

Страница 158: ...cont d 3 On the Service Tab select the Destination Device My Computer Select Service Type to Add select USB Quick Save then Add and then Save 4 Assign Network Storage to USB Quick Save under Destinat...

Страница 159: ...structions for setting up Network Storage on LOGIQ S8 R4 cont d 5 Select Button Tab Select a Print Button then add the USB Quick Save to the Printflow View by selecting If Copy to Dataflow is in Print...

Страница 160: ...tes for InSite Setup 3 82 3 13 4 Configuration Steps 3 83 3 13 5 Configuring InSite 3 83 3 13 8 Configuring Request for Service RFS 3 90 3 13 9 Verifying InSite Connection CSD 3 91 3 13 10 Verifying I...

Страница 161: ...e configuring the agent be sure that you have a network connection and check that it is configured You will not be able to perform checkout with the default network configuration 3 13 6 Configuring Ag...

Страница 162: ...te Connection R1 to R4 3 13 6 Configuring Agent Configuration cont d Figure 3 52 Configuration Home Page 4 In the Display Name field enter a descriptive name that is easy to identify with this device...

Страница 163: ...nt d 6 Configure the following mandatory fields as desired Continent Country City State Institution The Agent Configuration Tool screen opens Figure 3 53 Completed Agent Configuration Tool screen Must...

Страница 164: ...change the Directory field 4 In the Filter field it is recommended to set it to zip so that the system will only upload zip files to the back office Example Alt D logs 5 If a proxy is needed select E...

Страница 165: ...k out confirm the Agent Configuration Tool screen has correct URL IP address information Make corrections and repeat 1 Select PILOT2 or PRODUCTION from the Enterprise Server drop down list The Pilot s...

Страница 166: ...er s address Latitude No Arbitrary Longitude No Arbitrary Institution Yes Arbitrary Enter Customer s info Department No Arbitrary Enter Customer s info Building No Arbitrary Enter Customer s info Floo...

Страница 167: ...Section 3 13 Setting up InSite Connection R1 to R4 3 89 Scheme No Consult institutions IT personnel for Proxy settings Proxy User No Consult institutions IT personnel for Proxy settings Password No C...

Страница 168: ...E The RFS screen opens 3 Select the Users tab Click on Add Users and type the site s Contact Name Manual and machine generated RFS requests are directed to this person Items with red asterisks are req...

Страница 169: ...der LOGIQ S8 is found under folder ROOT CS_ULS_LOGIQ_S8_03 0 R1 x x ROOT CS_ULS_LOGIQ_S8_03 5 R2 x x R3 x x ROOT CS_ULS_LOGIQ_E8_03 5 LOGIQ E8 2 Request the OLC to attempt to connect to the system via...

Страница 170: ...oke Disruptive Mode and VCO by right click on the GE InSite icon at the bottom of the display and select Connect Clinical Lifeline Figure 3 59 Select Connect Clinical Lifeline 2 Confirm with the OLC t...

Страница 171: ...change to a new Device Type Contact the On Line center to seek help for clearing up the quarantine status 3 14 2 Verifying back office connection CSD 1 Login to FFA Site from Remote PC using Chrome we...

Страница 172: ...tically denied 1 On Console invoke Disruptive Mode and VCO by left click GE ICON and select Connect Clinical Lifeline 2 Login to FFA Site from Remote PC using Chrome web browser As of 2018 FFA Site ad...

Страница 173: ...ices Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept LOGIQ S8 SN Type Floor Room REV Comments CONTACT INFORMATION Name Title...

Страница 174: ...d Figure 3 61 Worksheet for DICOM Network Information LOGIQ S8 Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Address...

Страница 175: ...t safety and user information is available during the operation and service of the complete system 3 16 1 Product Locator Installation Card NOTE The Product Locator Installation Card shown may not be...

Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...

Страница 177: ...4 4 Functional Checks 4 18 4 5 Application Turnover Check List 4 48 4 6 Power supply test and adjustments 4 48 4 7 Site Log 4 49 Section 4 2 Required equipment An empty blank CD R DVD R and or extern...

Страница 178: ...e and cooling purposes NOTICE Energy Control and Power Lockout for LOGIQ S8 When servicing parts of the system where there is exposure to voltage greater than 30 Volts 1 Complete shutdown the system 2...

Страница 179: ...e the scanner During a normal boot you may observe Power is distributed to Peripherals Operator Panel control panel Monitor Front End and Back End Processor The Back End Processor and rest of the scan...

Страница 180: ...ou are acknowledging that you are authorized to do so and are trained in appropriate use and regulatory guidelines Figure 4 2 Login Window Example of R4 2 5x After you have established your password f...

Страница 181: ...Shutdown Use this button to shut down the system The entire system will shut down The power button turns on blue indicated Standby mode refer to Figure 4 1 Power Button on page 4 3 For the system with...

Страница 182: ...onds before turning it on again The system may not be able to boot if power is recycled too quickly NOTICE The The system will go to Power Assistant Mode Optional R2 or later if above procedures are s...

Страница 183: ...avoid injury or damage make sure nothing is within the range of motion before moving the monitor and monitor arm This includes both objects and people 19 LCD Monitor 23 Wide Monitor 22 Wide Monitor P...

Страница 184: ...dures 4 3 5 2 Monitor Arm Lock 1 Unlock the monitor arm 2 Verify the monitor can move up down vertically and left right horizontally 3 Align the monitor arm along center line of the system 4 Engage th...

Страница 185: ...operator panel while holding lever 3 Verify the operator panel is free to up down 4 Release the elevation lock lever disengage 5 Verify the operator panel does not up down CAUTION To avoid injury or...

Страница 186: ...vel Lock Lever R4 system 1 Hold Swivel Lock lever under the operator panel 2 Rotate the operator panel while holding lever 3 Verify the operator panel is free to rotate 4 Release the swivel lock lever...

Страница 187: ...otate wheel until it locks only one lock position in 360 deg turn 3 Verify Swivel Lock engages 4 Press b downward see Figure 4 9 5 Verify Brake is applied to the wheel i e Wheel does not rotate 6 Pres...

Страница 188: ...igure 4 10 Total Lock Casters 1 Press a see Figure 4 10 2 Rotate wheel until it locks several locations where caster engages 3 Verify Swivel Lock engages and Brake is ON 4 Press b see Figure 4 10 5 Ve...

Страница 189: ...hat the control panel the monitor arm lock firmly and flip down the monitor to prevent system damage CAUTION DO NOT place probes or the footswitch into the side tray when moving transporting the syste...

Страница 190: ...7 Power On Off on page 3 17 and remove the plug from the wall 4 All cables from off board peripheral devices must be disconnected from the console 5 Disconnect the footswitch from the console 6 Wind...

Страница 191: ...hile moving Take extra care when moving the system long distances and on inclines 5 degrees Ask for help if necessary DO NOT attempt to move the console using any cables or fixtures such as the probe...

Страница 192: ...Prevent vibration damage by driving cautiously Avoid unpaved roads excessive speeds and erratic stops or starts Load the system onto a vehicle 1 Load and unload the system to a vehicle parked on a lev...

Страница 193: ...tion lock the wheel brakes 2 Follow the set up procedure described in Section 3 6 Completing the Set Up on page 3 12 4 3 9 Service Platform Confirmation R1 to R4 1 From the touch panel select Utility...

Страница 194: ...4 6 B Mode Checks 4 23 4 4 7 M Mode Checks 4 25 4 4 8 Color Flow Mode Checks 4 27 4 4 9 PW CW Doppler Mode Checks 4 29 4 4 10 Tissue Velocity Imaging TVI Checks 4 31 4 4 11 SWE Shear Wave Shear Elasto...

Страница 195: ...R2 and R3 1 Power On Off 2 A N Keyboard 3 Measurement Selection Menu 4 Rotary and Joystick controls 5 TGC 6 Reverse key 7 User Define keys 8 BT Keys 9 Mode Gain XYZ Controls 10 Trackball Trackball Ke...

Страница 196: ...ion Arrow 11 F3 Eject media 12 F4 Activates DICOM Job Spooler screen 13 F5 Creates a Fast Key 14 F6 Plays a Fast Key 15 F7 Home Set Home 16 F8 Text1 Text2 17 F9 Grab last 18 F10 Word delete 19 Tab key...

Страница 197: ...e keys on the trackball area as described below The Trackball area consist of The Trackball used as a cursor control in acquisition mode scrolling control in freeze and as a selection tool like a mous...

Страница 198: ...Healthcare recommends the RMI 430GS phantom optional but it is not required It is the most current phantom optional recommended to our field service personnel and provides the necessary targets and ex...

Страница 199: ...ction 4 4 Functional Checks 4 23 4 4 6 B Mode Checks For information on the system s different modes as well as expected results refer to the latest revision of the Basic User Manual Chapter 5 4 4 6 1...

Страница 200: ...l Use Depth to adjust the range to be imaged Use Focus to center the focal point s around the region of interest Use Frequency move to higher frequencies or Frame rate move to lower frame rate to incr...

Страница 201: ...Dynamic Range Controls how echo intensities are converted to shades of gray thereby increasing the adjustable range of contrast Focus Number and Position Increases the number of transmit focal zones...

Страница 202: ...to adjust the range to be imaged Use the Frequency move to higher frequencies or the Frame rate control move to lower frame rate to increase resolution in image Use the Frequency move to lower frequen...

Страница 203: ...t one of the probes to the scanner s active probe connector see 3 6 4 Connecting Probes on page 3 15 for info about connecting the probes For available probes see Section 9 13 Probes on page 9 59 2 Tu...

Страница 204: ...a low setting for better flow resolution or a higher setting to more easily locate disturbed flows Adjust Frequency to optimize the color flow display Higher settings improve resolution Lower setting...

Страница 205: ...bring up a B Mode picture for navigation purpose and then add Doppler For information refer to the latest revision of the Basic User Manual Chapter 5 4 4 9 1 Preparations 1 Connect one of the probes...

Страница 206: ...r setting will improve resolution and the lower setting will increase the depth penetration Adjust Frame rate to a higher setting to improve motion detection or to a lower setting to improve resolutio...

Страница 207: ...solution 4 4 10 2 Preparations 1 Connect one of the probes to the scanner s left most probe connector See 3 6 4 Connecting Probes on page 3 15 for info about connecting the probes For available probes...

Страница 208: ...Check Select a probe 9L D and C1 6D Select ELASTO button In the Touch Panel select Shear Wave button if not already selected Verify that the Shear Wave button is available 4 4 12 Cineloop Check 4 4 1...

Страница 209: ...o worksheets press Clear 4 4 13 2 Check Circumference Area Ellipse Measurement 1 Press Measure once to display an active caliper 2 Move the Trackball to position the active caliper 3 Press Set to fix...

Страница 210: ...Q S8 LOGIQ E8 SERVICE MANUAL 4 34 Section 4 4 Functional Checks 4 4 13 4 Report Pages Measurements Calculations that are included on the worksheet can also be displayed on Report Pages Report Pages ca...

Страница 211: ...panel 1 Power ON the system 2 Check backlight on AN Keyboard B Button Freeze Key and TGC Knob 3 Select UTILITY in touch panel 4 Turn joystick knobs to adjust the backlighting of Touch Panel Light Keyb...

Страница 212: ...B image gain change 11 Press COMMENT key 12 Type in arbitrary strings from A N Keyboard Verify keyboard is functional 4 4 14 2 DVD Drive Test 1 Insert Blank CD R or DVD R Media into Optical Drive 2 Se...

Страница 213: ...dult And start scan 2 Verify calibration normally finished 3 Verify that B mode image appear CAUTION TAKE THE FOLLOWING PRECAUTIONS WITH THE PROBE CABLES KEEP AWAY FROM THE WHEELS DO NOT BEND DO NOT C...

Страница 214: ...splayed select Utility Application and Check Show ECG Tab NOTE Without the ECG option the ECG touch panel is not displayed NOTE For more information about ECG refer to the latest revision of the LOGIQ...

Страница 215: ...ity System Peripherals and confirm DVD is set as DVR media If not set it and restart the system 3 Insert the media Go to DVR Touch Panel menu 4 Start Recording Perform image scanning including Doppler...

Страница 216: ...er Test 1 Insert Gel Bottle into Gel Warmer 2 Power ON the warmer Figure 4 29 Gel warmer ON OFF switch 3 Verify Gel gets warm NOTE Depending on environment Gel requires 30 minute or more to warm up 4...

Страница 217: ...Check 4 4 18 1 Physical Abnormalities Check for no loose or cracked parts on above locations Figure 4 30 LOGIQ S8 System Exterior example No Parts name 1 Main monitor 2 Monitor Arm 3 Control Panel 4 R...

Страница 218: ...ction 4 4 Functional Checks Figure 4 31 LOGIQ E8 System Exterior No Parts name 1 Main Monitor 2 Monitor Arm 3 Control Panel 4 Rear Handle 5 Top Cover 6 Front Cover 7 Side Covers 8 Base 9 Casters 10 Re...

Страница 219: ...4 43 4 4 18 2 Label Check Figure 4 32 Label location R1 R3 system Console No Manufacturing date GE ULTRASOUND KOREA Ltd 65 1 SANGDAEWON DONG JUNGWON KU SEONGNAM SI GYEONGGI DO KOREA Serial No 100 120...

Страница 220: ...Check presence of Caution Label on the rear panel of the monitor 2 3 2 Check presence of Multi Caution Label 2 2 3 Check presence of Rating Plate 2 1 NOTE Rating Plate design slightly differ depending...

Страница 221: ...3 Check presence of Rating Plate 2 1 NOTE Rating Plate design slightly differ depending on Console destination At minimum simply check Console number manufacturing date and serial Number NOTE Some cou...

Страница 222: ...d Battery Diag can be performed with over 15 charging condition And actual battery kit is shipped with 20 30 charged condition so basically you don t need to check capacity for function test of extend...

Страница 223: ...nd white on FibroScan Module front panel are lit Figure 4 35 LED on FIbroScan 3 Select the assigned key to start FibroScan mode Screen will be dual screen FibroScan view will be showed on right side o...

Страница 224: ...st Refer to 7 9 3 System on page 7 130 4 6 2 Power Supply Adjustment There are no adjustments on the power supplies The DC Power is self regulated If a voltage is outside the specified range it means...

Страница 225: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 4 7 Site Log 4 49 Section 4 7 Site Log Table 4 5 Site Log DATE SERVICE PERSON PROBLEM COMMENTS...

Страница 226: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 4 50 Section 4 7 Site Log Table 4 5 Site Log Continued DATE SERVICE PERSON PROBLEM COMMENTS...

Страница 227: ...r Distribution scheme and probes NOTE When not otherwise specified the contents of this manual and reference to LOGIQ S8 applies to all LOGIQ S8 LOGIQ E8 LOGIQ S8 Vet models 5 1 2 Contents in this cha...

Страница 228: ...eneral Information LOGIQ S8 is a digital beam forming curved linear and phased array ultrasound imaging system The system supports various operating mode such as B Mode M Mode Color Flow CFM Power Dop...

Страница 229: ...Printer or V Nav 9 Color printer or Drawer 10 Probe port 4 active probe port and 1 parking port LE8 has the cover 11 Foot rest 12 Side tray option 13 Touch panel 14 CW pencil probe port 15 ECG Connect...

Страница 230: ...ide Monitor 2 Control panel 3 Probe holder 4 Gel warmer 5 Audio speaker 6 DVD Drive 7 Blind cover B W printer Color Printer Drawer V Nav controller or FibroScan module 8 Probe port 4 active probe port...

Страница 231: ...r Panel Figure 5 4 R4 and later Operator Panel Integrated A N Keyboard Interchangeable FREEZE and P1 Key User Define Keys BT Key Power Button Probe Holder Gel Warmer Probe Holder Option Probe Holder I...

Страница 232: ...be connector 5 2 2 System Options Figure 5 6 System Options DLP Probe Connector 4 ports Parking DLP Connector Single CW Connector CWD Option LOGIQ S8 only Optional Probe Holder ECG Connector Cabinet f...

Страница 233: ...SERVICE MANUAL Section 5 2 General Information 5 7 5 2 3 USB Ports Figure 5 7 USB Ports on the system Figure 5 8 OPIO USB port R4 system 23 Wide monitor Left side 22 Wide monitor Right side One direc...

Страница 234: ...d limit 5 2 3 3 USB ports on the monitor R3 and later The two USB2 0 Ports are used for USB Storage USB HDD and USB Flash Device NOTE USB ports on the monitor are non isolated and not fit to connect e...

Страница 235: ...8 SERVICE MANUAL Section 5 2 General Information 5 9 5 2 3 4 USB ports on the OPIO R4 or later The two USB3 0 Ports are used for USB Storage USB HDD and USB Flash Device Table 5 4 Specification Softwa...

Страница 236: ...2 Ethernet 1000 Base T 100 Base T 10 Base T 5 2 4 3 Audio MiniJack Normal Audio L R GND pin assign Line out 5 2 4 4 AC Inlet Figure 5 9 AC Inlet 1 USB Port USB2 0 Full Speed R3 and later No isolated...

Страница 237: ...button on the Operator Panel 3 BEP and system power up Sequence 1 Circuit breaker ON 2 AC DC Unit creates Stand by Power 3 Stby Power routed to COM Express 4 Stby Power routed to OPIO LED 5 Main ON s...

Страница 238: ...description A power loss may be due to The Mains Switch has been switched to OFF The Mains cable has been disconnected Brown out If a power loss occur all AC power distribution within the unit is los...

Страница 239: ...sed on LOGIQ S8 You can also refer to the Basic User Manual for more information 5 5 2 Contents in this section 5 5 1 Purpose of this section 5 13 5 5 2 Contents in this section 5 13 5 5 3 LOGIQ S8 so...

Страница 240: ...ference and enclosed area Yes Yes Yes Yes Yes Coded Contrast Provides contrast imaging capability Yes Yes Yes Yes Yes CW Doppler LS8 only Allows examination of blood flow data all along the Doppler Mo...

Страница 241: ...er a Static 3D or Real Time 4D acquisition N A Yes Yes Yes Yes VCI Static VCI Volume Contrast Imaging allows you to sweep smaller slices of data with a higher volume rate The resulting image shows an...

Страница 242: ...approach for clinical assessment of the fetal heart STIC offers an easy to use technique to acquire data from the fetal heart and to aid in visualization with both two dimensional and three dimension...

Страница 243: ...S8 Power Assistant hardware Y Y LS8 R4 Power Assistant hardware Y Y Battery for Power Assistant Y Y Y Y Extended battery for scan Y Y LS8 R3 CW Doppler kit Y Y Y FibroScan Module Y Y Peripheral Device...

Страница 244: ...V Nav eTRAX 18 20g starter kit Y Y Y Y V Nav eTRAX starter kit 12GA Y Y Y Y V Nav eTRAX starter kit 14GA Y Y Y Y V Nav Active Tracker kit Y Y Y V Nav Active Tracker kit MR Y Y Y Cardiac and ECG Optio...

Страница 245: ...re as UP D897 compati Mode Color Sony UP D25 Y Y Y Y Y LS8 Sony UP D55 Digital Color A5 Printer Y Y Y Y Color Sony UP DR80 Y Y Y Y UP DR80 can be used on LOGIQ S8 R2 2 1 or later Color Inkjet Printer...

Страница 246: ...SERVICE MANUAL 5 20 Section 5 7 Hardware Options LS8 UVC Off board installation kit Y Y Y Y LS8 Off Board UVC AC Adaptor Y Y Y Y Japan model is different H Cat Table 5 6 Hardware option list Hardware...

Страница 247: ...94227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 5 7 Hardware Options 5 21 5 7 1 LOGIQ S8 R4 FibroScan Option 5 7 1 1 FibroScan Screen Figure 5 10 FibroScan Screen 5 7 1 2 FibroScan Module Box Figure...

Страница 248: ...ptions 5 7 1 3 FibroScan mounting position Figure 5 12 FibroScan mounting position 5 7 1 4 FibroScan Signal Connection Figure 5 13 Signal Connection 5 7 1 5 FibroScan Probe Figure 5 14 FibroScan Probe...

Страница 249: ...activate XL probe Note If only one FibroScan probe is connected the exam type is automatically activated when entering the FibroScan mode CAP Activate the CAP measurement when the CAP option is insta...

Страница 250: ...Calibration Calibration Indicator Probe needs to be calibrated every 2 years A message will be displayed 14 days prior calibration expiration date as well as if the calibration has expired When perfo...

Страница 251: ...on Set South Africa Destination Set Argentina Destination Set Israel Destination Set Switzerland Destination Set Denmark Destination Set US Destination Set Japan Destination Set Australia New Zealand...

Страница 252: ...n 5 9 1 Air Flow Distribution Through the filter grid on the front of the system air flow into the LOGIQ S8 By means of the 1 FAN air is blown through the nest box and the warm air exits the system th...

Страница 253: ...ference The online versions of the operator manuals are available via the Help function F1 key on LOGIQ S8 s operator panel The translated user manuals are available on a CD ROM delivered with the sys...

Страница 254: ...for LOGIQ E8 This universal video converter needs a fan assy for cooling 4 LOGIQ E8 SW Option LOGIQ E8 does use LOGIQ S8 application software LOGIQ E8 SW is required to boot system as LOGIQ E8 When L...

Страница 255: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 5 11 LOGIQ E8 5 29 Figure 5 19 LOGIQ E8 Emblem position R4 system...

Страница 256: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 5 30 Section 5 11 LOGIQ E8 This page was intentionally left blank...

Страница 257: ...his chapter This chapter describes how to test and adjust the mechanical capabilities of the sytsem that may be out of specification Although some tests may be optional they should only be performed b...

Страница 258: ...adjusted values will be lost after loss of power Figure 6 1 Monitor Adjustment buttons NOTE 1280 x 1024 60Hz Table 6 1 Contents in this section Section Number Main Menu Sub Menu Range Setting for LOGI...

Страница 259: ...le The lowest level of black should just disappear into the background and the highest white should be bright but not saturated 1 Adjust the BRIGHTNESS by pressing the LEFT or RIGHT button to decrease...

Страница 260: ...move the cursor and the MENU middle key 3 Select 2 2 or 2 4 by pressing the MENU middle key 6 2 1 1 4 Mode 1 Press the MENU middle key of the monitor controls over 10 sec 2 Select the Picture Color T...

Страница 261: ...UNCTION Scale by pressing the LEFT or RIGHT button to move the cursor and the MENU middle key 3 Select FULL 5 4 NATIVE by pressing the LEFT or RIGHT button to decrease increase the value and the MENU...

Страница 262: ...to move the cursor and the MENU middle key NOTE MEMORY RECALL is Factory default If selecting MEMORY RECALL all settings will be back to factory default status 6 2 1 2 4 SBC 1 Press the MENU middle k...

Страница 263: ...SITION by pressing the LEFT or RIGHT button to decrease increase cursor and the MENU middle key 3 Adjust the H POSITION by pressing the LEFT or RIGHT button to decrease increase the value default 50 6...

Страница 264: ...touch panel 2 Adjust Room profile Color Profile Gamma Brightness Contrast Room profile User defined and Color Space 22 OLED monitor only as necessary Figure 6 7 Utility second page 22 OLED Monitor exa...

Страница 265: ...main display If User Defined is selected the user can adjust Brightness Contrast by Brightness Contrast control at the bottom of the Touch Panel NOTE Brightness Contrast setting is displayed only when...

Страница 266: ...n This procedure is a means to do a system specific calibration correction Note that the field calibration only applies to the specific type of probe 9L D ML6 15 D C1 5 D e g that is used Therefore it...

Страница 267: ...e 2nd Environmental Quality Indicator in the upper left corner of the screen to initiate a mouse over window 3 Move the transducer across the surfaces of the phantom optional where you will be scannin...

Страница 268: ...ng the probe steady move the Windows Pointer over the point of interest and press the Right Trackball key A GPS Target Marker labeled 1 is placed on the point of interest 5 Change the orientation on t...

Страница 269: ...est identified in step 5 For example turn the probe 90 degrees or move the probe to a different surface of the phantom optional The GPS marker will help to guide back to the same point of interest mak...

Страница 270: ...series of calculations are performed in 5 steps Once the fifth step is complete a value appears on the status bar This message appears briefly so watch for it If you miss this message press Calibrati...

Страница 271: ...trouble images and system data logs for acquisition through remote diagnostics or to be sent back to the manufacturer for analysis There are different options to acquire this data that would give dif...

Страница 272: ...stem Remote access to the LOGIQ S8 requires permission and customer input to run diagnostics 7 2 3 Local Access R1 to R4 1 If not already in read mode freeze the image 2 Select Utility on the Touch Pa...

Страница 273: ...access to the LOGIQ S8 scanner requires permission and customer input before a GE service technician or OLC can access the customer s ultrasound scanner remotely Disruptive Mode can be selected by th...

Страница 274: ...ote access to your LOGIQ S8 scanner following InSite Notification appears on the systems screen 1 Enable Disruptive Mode feature by confirming Yes NOTE If the customer does not wish to have diagnostic...

Страница 275: ...ce or through remote diagnostics InSite These data can be used to perform service at the device or can be sent back to the manufacturer for analysis 7 3 2 Contents in this section 7 3 1 Purpose of thi...

Страница 276: ...r being returned to the manufacturer NOTE This information is normally collected with the Alt D or Gather Logs utility Product Name LOGIQ S8 or LOGIQ E8 1 From the touch panel select Utility System Ab...

Страница 277: ...ointer on the GE InSite icon at the bottom of the display 2 Press the right Trackball Set key Select Contact GE The RFS screen opens Figure 7 7 Contact GE Request for Service NOTICE NOTICE Service Con...

Страница 278: ...onfirmation screen is displayed Figure 7 8 Request submitted 3 Write down and keep the Reference Number for follow up procedures then click OK NOTE f the service platform is not configured an Error me...

Страница 279: ...ects a screen capture of the image monitor user defined presets and the following logs Keyboard Shadow Log Error Logs Crash Log Vital Product Data DICOM Logs Windows Event Logs Diagnostic Logs Service...

Страница 280: ...or CD DVD the system will automatically format if you insert an unformatted disk gather logs and write it out to the disk The subsequent file is compressed and time stamped The screen capture is a bit...

Страница 281: ...or later Zip file configuration 7 3 4 1 Marking Log files If a customer is experiencing issues during operations the event can be marked and logged by pressing Alt 1 or Alt 2 when they occur When Alt...

Страница 282: ...LOGIQ S8 LOGIQ E8 SERVICE MANUAL 7 12 Section 7 3 Gathering Troubleshooting Data R1 to R4 7 3 5 Capturing Service Logs with ALT D The following is a list of the Service logs captured during an ALT D l...

Страница 283: ...s that can then be viewed by the OLC 7 3 6 2 To Capture a Screen Image Using the Shortcut With the desired image displayed on the screen press Ctrl and PrtSc print screen keys simultaneously If you wa...

Страница 284: ...under d log Sniffer folder to ensure they form part of general log collection Alt D or Collect Log 1 Press Alt N 2 On the network sniffer screen select Capture Options Figure 7 13 Network Sniffer Scre...

Страница 285: ...rk Sniffer cont d Use filters to limit the captured data Filter by IP or by IP and port number By IP only press on Capture Filter 3 Select IP only Use the IP address of the system or the one from the...

Страница 286: ...or example DICOM port 104 6 Enter the string with the following syntax port xxx and host yy yy yy yy where xxx is the port number of My Computer in the LOGIQ S8 and yy yy yy yy is the IP address of th...

Страница 287: ...y A Packet List pane the packet list pane displays all the packets in the current capture file Each line in the packet list corresponds to one packet in the capture file B Packet Details pane shows th...

Страница 288: ...11 Type dcm and press Apply The display should filter all DICOM packets filtering out image data Figure 7 17 Filter Tool Display 12 Select Capture Stop Or select the icon in the task bar to stop the...

Страница 289: ...rather than the entire capture 15 Select SAVE If you perform Alt D or Gather Logs these sniffer logs will be included in the zip file Figure 7 19 Select Capture Display 16 Exit the sniffer application...

Страница 290: ...NOTE When using the Common Service Desktop do NOT minimize any of the Common Service Desktop windows If you minimize them they end up in the lower left corner of the screen behind the Service Desktop...

Страница 291: ...vels Each user is only granted access to the tools that are authorized for their use Every access request whether successful or not will be logged into a service access log that is viewable to authori...

Страница 292: ...s etc 2 Connected Probes Connected and System recognized probe listed 3 Options Installed Enabled Options and its expiration date 4 Windows Printers Connected and system recognized peripheral listed 5...

Страница 293: ...bles to view a number of different system logs 1 LOGS enables to view following logs System Displays all the system logs including errors and additional details Power Displays all the power logs inclu...

Страница 294: ...interface to provide access to common service components and perform diagnostics Non Interactive The tests are performed without the user s intervention Interactive The user is required to perform an...

Страница 295: ...4 Service Desktop CSD R1 to R4 7 25 7 4 7 Image Quality In the Image Quality page you can verify image quality NOTE This page is not populated in this version 7 4 8 Calibration There is no accessible...

Страница 296: ...n the Configuration page you can view and modify different device informations and configurations in the InsiteExC Agent Configuration option field Figure 7 25 Common Service Desktop Configuration NOT...

Страница 297: ...Application link an event log relative to application events System link an event log relative to system events Log Name enter the Log Name you want to view and click the View button Figure 7 26 Event...

Страница 298: ...12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 7 28 Section 7 4 Service Desktop CSD R1 to R4 Figure 7 27 Disruptive Mode Window 7 4 10 3 Disk Usage View capacity and usage statistics for the different disk drive...

Страница 299: ...MANUAL Section 7 4 Service Desktop CSD R1 to R4 7 29 7 4 10 4 IP Configuration View Windows IP configuration and LAN connection data Figure 7 29 IP Configuration 7 4 10 5 Network Status View data for...

Страница 300: ...View the Windows Services that are started and running A Windows Service is a computer program that has been automatically started and is running in the background on the computer Figure 7 31 Windows...

Страница 301: ...irus Software Control Select Disable to disable the Anti Virus software if needed Figure 7 33 Anti Virus Software Control Window NOTE R4 If the Advanced Security software option installed and the Anti...

Страница 302: ...s system Figure 7 34 Shared Resources Window 7 4 10 10 System Shutdown System Shutdown gives you the ability to Restart or Shutdown the system when using Virtual Console Observation from a remote comp...

Страница 303: ...an be gathered using the Gather Logs shortcut on the Windows desk top Figure 7 36 Gather Logs Utility WIndow In R4 and later Gather Logs Utility will not be collecting logs that contain protected info...

Страница 304: ...images available at this time Compress Files compresses images into a zip file Delete Files deletes the images from the image Export Service directory Figure 7 38 Image Compress Delete Utility Window...

Страница 305: ...MANUAL Section 7 4 Service Desktop CSD R1 to R4 7 35 7 4 11 Utilities Scanner Utilities 7 4 11 1 Reset Database Remove Patient data from database Figure 7 40 7 4 11 2 Clean Userdefs Remove User defs...

Страница 306: ...verify DICOM connectivity between the scanner and DICOM devices on the network 1 Enter AE Title IP Address and Port values of the DICOM device 2 Check the Loop checkbox to repeat the operation or lea...

Страница 307: ...DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 7 4 Service Desktop CSD R1 to R4 7 37 7 4 12 Replacement Field is not populated 7 4 13 PM Field is not populated Figure 7 43 Common Servi...

Страница 308: ...athering log files To gather log files the system requires a user to be logged in The error message below appears when hitting Alt D key without any user logged in 7 5 2 Collecting vital system inform...

Страница 309: ...e flagging issues during operation Alt F10 Displays SW Shut Down If the LOGIQ S8 can be shut down by using these keys it means that SW is working and the problem could be the ON OFF switch or the wire...

Страница 310: ...Alt D is available at all times and collects a screen capture of the image monitor user defined presets and these logs Keyboard Shadow Log protected information Error Logs Crash Log protected informa...

Страница 311: ...g_ SN _ DATE _ TIME _DB zip log_ SN _ DATE _ TIME _ProtectedInfo_DB zip Where SN is the serial number DATE is the date in format YYMMDD and TIME is the time in format HHMMSS 4 To include a duration ch...

Страница 312: ...ains the file An example is shown in Figure 7 45 Example of Zipped Trouble Image and Logs File on page 7 42 Figure 7 45 Example of Zipped Trouble Image and Logs File Figure 7 46 Contents of the Zipped...

Страница 313: ...t information to GE under ADMIN System Admin To collect the image press the P1 key copy to hard drive This places an unidentified jpeg image in the D Log DL folder This folder will be transferred to t...

Страница 314: ...e other network sniffer to analyze it To troubleshoot network capture logs 1 On the LOGIQ S8 run the Network Capture utility to generate etl and cab files for analysis For more information see Network...

Страница 315: ...45 6 To export the etl file to cab select File Save As and then select Save As or Export If the network capture was of a wireless connection Microsoft Message Analyzer creates a pcap file that does no...

Страница 316: ...figuration 7 66 7 6 1 Purpose of this section This section describes the features of the Service desktop These are the different levels of access to the Service desktop Service Basic access Class A a...

Страница 317: ...LOGIQ S8 may be limited When Disruptive mode is Off some service functionality on the Service desktop is not available and user operation of the LOGIQ S8 is normal Additionally the ability to enable D...

Страница 318: ...Section 7 6 Service Desktop R4 2 5x and later 7 6 4 Licenses With Service Basic Access Class A these are the available options HOME Utilities Change Password Data Transfer Delete Files Gather Logs Net...

Страница 319: ...7 6 Service Desktop R4 2 5x and later 7 49 7 6 5 Home Home configurations vary depending upon the purchased service level Figure 7 49 Home with Class A Access For more information see System Informat...

Страница 320: ...Q S8 When the LOGIQ S8 has been successfully configured with the back office these elements will have the corresponding values Agent Registered will be Yes Agent Quarantine will be No Agent CRM Verifi...

Страница 321: ...igured Agent Quarantine Quarantine status of the agent Valid values are Yes The agent has more than one device registered with the same CRM Number in the back office No The agent has one device regist...

Страница 322: ...7 5 Software Status Element Description System Date Current date in the format day month date year System Time Local time based on the last time the system desktop was refreshed in the format hh mm ss...

Страница 323: ...shows all the elements available on Connected Probes with descriptions For more information see Home on page 7 49 Table 7 6 Connected Probes Element Description Active Probe Temperature Celsius When a...

Страница 324: ...later 7 6 6 Utilities Utilities configurations vary depending upon the service class For more information see Change Password on page 7 55 Data Transfer on page 7 57 Delete Files on page 7 59 Gather L...

Страница 325: ...elements available on Change Password with descriptions CAUTION RISK OF LOSING DATA IF THE PASSWORD IS LOST GE WILL NOT BE ABLE TO RECOVER OR RESET IT LOSS OF A PASSWORD MAY RESULT IN THE LOSS OF PAT...

Страница 326: ...f the password is changed when offline the back office will not be updated and you will have to change the password again with the LOGIQ S8 online There is no notification when this password has been...

Страница 327: ...descriptions Table 7 8 Data Transfer Element Description Type of Upload Type of log file For example Incremental Logs or Full Logs Monitoring Logs System Logs and Windows Logs are incrementally transf...

Страница 328: ...hese settings Transfer Machine Logs to GE Default is checked Transfer Operator Information to GE Transfer Images captured without patient information to GE Default is checked 3 Click Save 4 Navigate t...

Страница 329: ...vice class licenses To access Delete Files select Utility second page Service Utilities Delete Files Figure 7 55 Delete Files This table shows all the elements available on Delete Files with descripti...

Страница 330: ...lable to all service class licenses To access Gather Logs select Utility second page Service Utilities Gather Logs Figure 7 56 Gather Logs This table shows all the elements available on Gather Logs wi...

Страница 331: ...e elements available on Network Capture with descriptions Table 7 11 Network Capture Element Description Network Capture Parameters Maximum Size Allowed size of the generated log file Valid value are...

Страница 332: ...devices are configured only All will be available 3 Select Start Network Capture to start the process This causes the network capture to start enables the Stop button and updates the Network Capture S...

Страница 333: ...SSA License This table shows all the elements available on SSA License with descriptions Table 7 12 SSA License Element Description Class M Key Details Class M Key Status Status of the SSA key Valid v...

Страница 334: ...Party Software Licenses displays the third party software licenses used as part of the service platform The information on Third Party Software Licenses is available to all service class licenses To...

Страница 335: ...e of the screen Press Show All to view all of the activated options The information on Options is available to all service class licenses To access Options navigate to Utility second page Service Opti...

Страница 336: ...Update contact details The information on Agent Configuration is available to all service class licenses To access Agent Configuration navigate to Utility second page Service Agent Configuration Figu...

Страница 337: ...same number as listed on the rating plate Serial Number Serial number of the agent read only If the agent is not registered with a serial number this field is populated with the serial number of the L...

Страница 338: ...onse rate between the LOGIQ S8 and a DICOM PACS server 7 82 7 7 20 When the touch panel does not response 7 83 7 7 21 Limitation message during AFI Analysis 7 83 7 7 22 Noise Troubleshooting 7 84 7 7...

Страница 339: ...Switch till it changes to green System starts booting Check cables between Console and Monitor Press Standby Switch for approximately 5 sec until system is down Is there any picture on the monitor Is...

Страница 340: ...ing Trees Instructions and Tech Tips 7 7 4 Noise disturbs the Image Figure 7 63 Noise disturbs the Image Troubleshooting Noise is shown in Image YES NO Is there any electrical devices that affect the...

Страница 341: ...sensitivity Trackball has low sensitivity Set up the appropriate trackball response time via Speed and Acceleration in the Utility System General Trackball Refer to Figure 3 23 on page 3 33 Is the tra...

Страница 342: ...lint free cloth e g lens cleaner cloth in accordance with procedure below 1 Remove the ball from the operation panel 2 Clean the crane ring and ball 3 Clean the aiming balls the protection covers ove...

Страница 343: ...ed with paper Load paper by using the correct format Check paper size small large etc Check paper tray to fit with the used paper format Is printing media ink cartridge etc empty Depending on the Prin...

Страница 344: ...rints do not match monitor Verify factory default settings Chapter 3 in the User Manual calls out suggested settings for various exam and lighting conditions Video test patterns are not clear bright p...

Страница 345: ...d be lower position ON For 22 and 23 wide monitor Switch No 8 should be lower position ON Fault symptom Check these items Cannot select probe from touch panel Make sure 4D option is installed Probe wi...

Страница 346: ...ugh if battery remaining capacity is quite low system automatically go to Power saving mode Cannot use Printer Cannot use Universal Video Converter This is system spec AC powered modules cannot be use...

Страница 347: ...mproper setting or dust in a cap Before replacing an encoder verify the setting dust of a cap NOTE It occasionally sounds like a correct setting from a cap when assembled it improperly 7 7 13 Sticky b...

Страница 348: ...ubleshooting Trees Instructions and Tech Tips 7 7 15 PC Printer Installation If the following message displays when you connect the PC Printer first time to the system execute the following procedures...

Страница 349: ...roScan probe is connected to FibroScan module Pull off the connected FibroScan probe from FibroScan module and reconnect it again Apply pressure to FibroScan probe with your finger tip If still not su...

Страница 350: ...int cursor Pointing cursor is available for only the following red area of the screen during FibroScan Mode 3 Measurements in Report Worksheet Order of the measurements isn t equal between FibroScan R...

Страница 351: ...g generated the button may not respond Example 2 Do not press the button on FibroScan probe when FibroScan Report is selected and the system starts generating the report WORKAROUND Do not press the bu...

Страница 352: ...und The system is accidentally turned off during the disk encryption is being configured Turn on the system The disk encryption configuration will be continued Lost the USB key to unlock the disk encr...

Страница 353: ...nse From R4 system the touch panel changes to electrostatic type If the keys does not respond clean the touch panel first If the gel or somehting remains on the touch panel it does not reposnd 7 7 21...

Страница 354: ...conditions when the noise appears Table 7 16 Noise Conditions The noise is present Determine If the test fails see All the time n a 7 7 22 2 Noise picked up from the air on page 7 85 and 7 7 22 3 Noi...

Страница 355: ...en the LOGIQ S8 is sharing a power line with other equipment 7 7 22 4 Intermittent noise Intermittent noise can be caused by other equipment Table 7 17 Noise Conditions Picked Up from the Air Symptom...

Страница 356: ...incoming power source to the LOGIQ S8 is only a conduit See Minimal floor plan suggestions on page 2 9 7 7 23 3 Different system Try another LOGIQ S8 at the same location and look for the same noise...

Страница 357: ...e network Media type not set to Auto Select Check the speed of your connection Media type should be set for Auto Select Remember that every time the system is re ghosted that setting goes back to the...

Страница 358: ...Cine or Secondary Capture Destination unable to view the particular photometric interpretation Check what compression and photometric interpretation the destination device supports and adjust if poss...

Страница 359: ...es not allow you to remove it You can uncheck the attribute and save but when returning to the same page the attribute remains checked See configuration screen on Utility Admin Users Workaround Remove...

Страница 360: ...DeletePatientData Allow to delete patient exam Can remove patients and exams if user has this right Y DeleteReport Allow to delete report But this right does not work any user can remove reports The...

Страница 361: ...down 1 Select OK and reboot the system 2 If the same message appears after reboot shut down the system and turn off the breaker Then turn on the system according to Power On Off on page 3 17 Cause Ha...

Страница 362: ...ower mode power saving locked keys etc No message Condition Plugged Warning Battery has deteriorated Replace the battery for proper function Condition Battery has deteriorated Shown periodically 14 da...

Страница 363: ...d the system on boot up Message Shutting down FibroScan Module please wait The system will be shutting down in a moment The system will shut down FibroScan Module safely before shutting down itself Th...

Страница 364: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 7 94 Section 7 8 Error messages This page was intentionally left blank...

Страница 365: ...12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Section 8 1 Overview 8 1 Chapter 8 Renewal Parts Section 8 1 Overview 8 1 1 Contents in this chapter 8 1 Overview 8 1 8 2 Optional Peripherals and Accessories 8 2 8...

Страница 366: ...Peripherals and Accessories Printers No Part Name Part Number Description Qty 1 Digital B W Video Printer Sony UP D897 5160458 Digital B W Video Printer USB Port 2 Digital B W Video Printer Sony UP D...

Страница 367: ...2 Optional Peripherals and Accessories 8 3 NOTE The illustrations may not correspond to the actual product 5 HP OfficeJet PRO 8100 ePrinter 5716062 6 HP OfficeJet PRO H8210 5763696 Table 8 1 Optional...

Страница 368: ...O ASO SRI HD Advanced3D Volume Navigation Biopsy Non RoHS 5499605 RoHS Compliant probe 5488219 C2 9VN D 5476030 5476031 Japan model 10C D Pediatric Neonatal Vascular No transcranial B CHI CF PDI M Ana...

Страница 369: ...Non RoHS Replaced by 5499594 5499594 RoHS Compliant probe GE 3MIX M5S D Cardiac Transcranial Abdomen B CHI CF PDI M MCF Anatomical M PW CW TVI TVD B Flow B Flow Color Contrast Virtual Convex LOGIQVie...

Страница 370: ...atomical M PW CW TVI TVD Virtual Convex ATO ASO SRI HD Note Need DLP2RN adapter to connect Note 6Tc RS probe requires special handling Refer to the TEE Probe User Manual enclosed with 6Tc RS probe Non...

Страница 371: ...ew ATO ASO SRI HD Advanced3D 5482856 L3 12 D Vascular No transcranial OB Small Parts Breast Musculoskeletal B CHI CF PDI M Anatomical M PW B Flow B Flow Color BF HD Color Contrast Strain Elastography...

Страница 372: ...tion 8 4 Power Cord Section 8 4 Power Cord Table 8 3 Power Cord No Part Name Part Number Description Qty 1 Power Cord for India 5182611 1 2 Power Cord for Switzerland 5182235 1 3 Power Cord for UK HK...

Страница 373: ...wer Cord for Argentina 5182942 1 8 Power Cord for Israel 5182453 1 9 Power Cord 110V for Canada USA JPN 2388982 Power Cord 110V for Canada USA JPN Power Cord 200V for Thailand 1 10 Power Cord 220V for...

Страница 374: ...995 2388982 1 13 Power Cord for Brazil 5399665 1 14 Power Cord for South Africa 5408183 1 15 Power Cord for Japan 5408490 1 16 Power Cord for Taiwan 5792418 1 17 Power Supply Cord EUROPE KOREA 10A 250...

Страница 375: ...apter This chapter describes Care Maintenance on the scanner and peripherals These procedures are intended to maintain the quality of the ultrasound systems performance Read this chapter completely an...

Страница 376: ...uality control testing must occur regularly The same tests are performed during each period so that changes can be monitored over time and effective corrective action can be taken Testing results corr...

Страница 377: ...y W Weekly M Monthly A Annually Table 9 1 Customer Care Schedule Item Service at Indicated Time D W M A Notes Air Filter Grid Remove the filter grid and clean the air filter Recommend to clean filter...

Страница 378: ...ctive maintenance or as required by your facilities QA program Peripheral Leakage Current Checks Also after corrective maintenance or as required by your facilities QA program Surface Probe Leakage Cu...

Страница 379: ...apter 3 Refreshing battery refresh the battery every 6 months for the user Refresh procedure 1 Turn on the system 2 Wait until the battery is fully charged it takes at least 1 hour to fully charge the...

Страница 380: ...Tool Comments Digital Volt Meter DVM Anti Static Kit Typically includes antistatic mat wrist strap and ground cable Anti Static Vacuum Cleaner If available on site Safety Analyzer Any calibrated Elec...

Страница 381: ...ion Certificate see Ultrasound INSPECTION CERTIFICATE on page 9 37 Note all probes and system options 3 Power up Turn the system power on and verify that all fans and peripherals turn on Watch the dis...

Страница 382: ...is mode of operation 5 3D Mode Where applicable verify basic 3D Mode operation Check the basic system controls that affect this mode of operation 6 Real Time 4D Mode Where applicable verify basic Real...

Страница 383: ...W video page printer Clean heads and covers if necessary 2 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 3 Media Verify hardcopy output of t...

Страница 384: ...a soft damp cloth Caution DO NOT spray any liquid directly onto the LOGIQ S8 covers monitor keyboard etc Table 9 8 Appropriate Cleaning Disinfectant Agents Cleaning Disinfectant Agent Monitor Glass Mo...

Страница 385: ...t cover of cabinet with hand and pull out the air filter Figure 9 1 Air filter location 2 Dust the filter with a vacuum cleaner and or wash it with a mild soapy solution If washed rinse and dry the fi...

Страница 386: ...the filter with a vacuum cleaner and or wash it with a mild soapy solution If washed rinse and dry the filter before re installation 4 Put back the air filter with both hands and the front cover Figur...

Страница 387: ...ck all wheels and casters for wear and verify operation of foot brake to stop the unit from moving and release mechanism Check all wheel locks and swivel locks for proper operation 7 Cables Connectors...

Страница 388: ...precautions can result in serious injury and equipment damage Failure to properly handle or maintain a probe may also void its warranty Figure 9 5 Probe handling icons Any evidence of wear indicates t...

Страница 389: ...re Card instructions supplied with each probe inside the transducer boxes for disinfectants and gels that are compatible with the surface material of the probes CAUTION CAUTION To help protect yoursel...

Страница 390: ...L 9 16 Section 9 7 Using a Phantom Section 9 7 Using a Phantom The Use of a Phantom is not required during Preventive Maintenance Customer may use it as part of their Quality Assurance Program tests R...

Страница 391: ...maged insulation and adequacy of strain relief and cable clamps Locate and examine all associated transducers Inspect the cables and strain relief at each end Inspect the transducer enclosure and lens...

Страница 392: ...ed service personnel or GE for further analysis Table 9 11 Leakage current limits for Ultrasound system operation on 100 120 Volt mains US Canada Japan Leakage Current Test System Power Lift Ground PE...

Страница 393: ...00 A CF 50 A Table 9 14 Equipment Type and Test Definitions Applied Parts AP Parts or accessories that contact the patient to perform their function For ultrasound equipment this includes transducers...

Страница 394: ...combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used Figure 9 6 Typical Alternate Outlet Tester NOTE No o...

Страница 395: ...nd wire resistance should be less than 0 2 ohms Reference the procedure in the IEC60601 1 Figure 9 7 Ground Continuity Test NOTE The Safety Analyzer can also be used instead of Ohmmeter Refer to the m...

Страница 396: ...s applicable Record the highest reading 1 Connect Safety analyzer to wall AC power outlet 2 Plug the equipment under test power cable into the receptacle on the panel of the meter 3 Connect the meter...

Страница 397: ...ach set of test conditions The actual location of the test probe may vary by Ultrasound system NOTE Record all data in the Electrical safety tests log NOTE Values in italics font are given as examples...

Страница 398: ...l AC power outlet 2 Plug the equipment under test power cable into the receptacle on the panel of the meter 3 Connect the ECG cable to the scanner and the Patient leads to the analyzer 4 Select the Pa...

Страница 399: ...MANUAL Section 9 8 Electrical Safety Tests 9 25 9 8 7 2 Generic Procedure cont d For each combination the operating controls such as the lead switch should be operated to find the worst case conditio...

Страница 400: ...ation under each set of test conditions specified in Record all data on the EQC inspection certificate Also known as Patient Auxiliary Current 1 Connect Safety analyzer to wall AC power outlet 2 Plug...

Страница 401: ...power outlet 2 Plug the equipment under test power cable into the receptacle on the panel of the meter 3 Connect the ECG cable to the scanner and the Patient leads to the analyzer 4 Select the Lead i...

Страница 402: ...configuration of the test equipment 9 8 10 1 Generic Procedure on Probe Leakage Current The most common method of measuring probe leakage is to partly immerse the probe into a saline bath while the p...

Страница 403: ...urrent path NOTE The Saline solution is a mixture of water and salt The salt adds free ions to the water making it conductive Normal saline solution is 0 9 salt and 99 1 water If ready mixed saline so...

Страница 404: ...as applicable NOTE Values in italics font are given as examples only Table 9 18 Typical data format for recording probe source leakage CAUTION Equipment damage possibility Never switch the Polarity or...

Страница 405: ...nnected from the Ultrasound system Apply mains voltage over the insulation barrier Between protective earth on the probe connector and an electrical anode in saline solution The patient applied part o...

Страница 406: ...test setup of non TEE Probes can be as illustrated in 9 8 10 Probe leakage current test on page 9 28 A typical test setup for TEE probes could be as indicated below Figure 9 14 TEE Probe Leakage Isola...

Страница 407: ...ng some other grounded surface 9 8 12 2 Generic Procedure on Leakage Current Refer to section 9 8 10 1 Generic Procedure on Probe Leakage Current for detail Put the e TRAX Needle sensor into Saline wa...

Страница 408: ...all data on the inspection certificate Table 9 19 Typical data format for recording e TRAX Sink leakage DANGER DANGER Electric Shock Hazard Line voltage is applied to e TRAX during this test To avoid...

Страница 409: ...s are normal from probe to probe Other variations will result from differences in line voltage and test lead placement The maximum allowable leakage current for body surface contact probe differs from...

Страница 410: ...9 When There s Too Much Leakage Current 9 9 5 ECG Fails Inspect cables for damage or poor connections 9 9 6 Ultrasound Equipment Quality Check EQC and IQC Download and use the latest version of these...

Страница 411: ...equency Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Num...

Страница 412: ...ing config System Ground Continuity Chassis Source Leakage Current Probe Chassis Source Leakage Current Wheel Chassis Source Leakage Current CRT Patient Lead Source Leakage Lead to Ground Patient Lead...

Страница 413: ...tance 1 31 D Dangerous Procedure Warnings 1 27 Dataflow Examples 3 67 Device Connectivity 3 76 Diagnostics 7 24 DICOM Adjust Timeout for DICOM Servers 7 88 DICOM Option Pre Installation Requirements 2...

Страница 414: ...Arrangement USA Canada 9 20 P Packing Materials Recycling Information 3 10 Periodic Maintenance 9 1 Schedule 9 3 Periodic Maintenance How often 9 3 Peripherals Connection 3 37 Physical Inspection 9 1...

Страница 415: ...te Requirements Mandatory 2 8 Sleep Mode 3 23 Sneaker Net 3 68 Software Option Installation 3 35 Sony UP D25MD Installation 3 58 Sony UP D897 Installation 3 49 Special Tools Supplies and Equipment 9 6...

Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...

Страница 417: ......

Страница 418: ...neral Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare GE Ultrasound Korea 65 1 Sangdaewon dong jungwon gu Seongnam si Gyeonggi do 462 120 Korea www gehe...