GE M
EDICAL
S
YSTEMS
D
IRECTION
5184024-100, R
EVISION
5
LOGIQ
WORKS
® B
ASIC
S
ERVICE
M
ANUAL
Chapter 3 Installation
3-145
•
Name (AMI format only):
Specify the name (for teleradiology purposes) of this system. This name
will be used to identify the system when it makes connections with other (LOGIQworksor
RadWorks) systems. Data received on other LOGIQworks machines (and sent from this machine)
will have their origin set to this name. The name is best kept short and concise.
•
Description (AMI format only):
Give a description of the machine you are working on. This is free
format text.
•
Maximum Retries:
Specify the maximum number of times a study will be re-sent to a destination
if it has previously failed. A new attempt is only made if the previous attempt was 'non-fatal' in
nature, such as a line interruption. If the number of attempts to transmit a study reaches the
specified maximum, the job will be set to an error state.
•
Retry Interval (sec.):
Specifies the (minimum) waiting time before a transmission (that failed on a
previous attempt) can be tried again.
•
Thread Priority:
Priority of the transmission thread within the Connection Service. Changing this
setting may influence system performance.
Figure 3-126 Connection Service Properties - Transmit (Default settings)