GE M
EDICAL
S
YSTEMS
D
IRECTION
5184024-100, R
EVISION
5
LOGIQ
WORKS
B
ASIC
S
ERVICE
M
ANUAL
3-2
Section 3-1 - Overview
3-1-3
Safety Reminders
WARNING
WARNING
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
WARNING
WARNING
Two people should unpack the unit because of its weight. Two people are required whenever a part
weighing 22 kg (50 lbs.) or more must be lifted.
CAUTION
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to its
operating environment.
Table 3-2
LOGIQworks Acclimate Time
°
C
-40
-35
-30
-25
-20
-15
-10
-5
0
5
10
15
20
25
30
35
40
45
50
55
60
°
F
-40
-31
-22
-13
-4
5
-50
41
32
41
50
59
68
77
86
95
104
113
122
131
140
hrs
20
18
16
14
12
10
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
WARNING
WARNING
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a
three to two prong adapter. This defeats safety grounding.
WARNING
WARNING
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQworks and