Chapter 3 Installation
3-141
•
Generic Receive
-
Verify Origin:
Check this check box if you want the system to perform 'origin checking'. This
option is available for DICOM transfers made to this system. When checked, the system will
verify the presence of the origin (i.e. the system sending data to this system) in the Origin
database.
This setting is also valid for receive activity in DICOM 3.0 format.
-
Auto Decompress Loss less Compressed Studies:
This setting controls whether studies
that are received in a loss less compressed format are automatically decompressed when
they are received. By default, this option is disabled. This will save storage space on the
receiving side, but may increase loading time when viewing the study. The reverse effect
(increased storage space, but decreased loading time) can be achieved by enabling this
option.