Chapter 3 Installation
3-155
-
Auto Decompress Loss less Compressed Studies:
This setting controls whether studies
that are received in a loss less compressed format are automatically decompressed when
they are received. By default, this option is disabled. This will save storage space on the
receiving side, but may increase loading time when viewing the study. The reverse effect
(increased storage space, but decreased loading time) can be achieved by enabling this
option.
-
Auto Decompress Lossy Compressed Studies:
This setting controls whether studies that
are received in a lossy compressed format are automatically decompressed when they are
received. By default, this option is disabled. This will save storage space on the receiving
side, but may increase loading time when viewing the study. The reverse effect (increased
storage space, but decreased loading time) can be achieved by enabling this option.
-
Auto Compress Studies:
Check the check box if you want LOGIQworks to automatically
compress incoming studies as they are received. This will save disk space, but performance
will be affected to some extent when studies are opened for viewing, since LOGIQworks will
have to decompress them before they are displayed. Select the type of compression you want
RadWorks to use from the Format drop-down list. If the format allows different compression/
quality settings (such as lossy JPEG), select a value from the Parameters drop-down list.
•
Success:
This parameter specifies the Windows.WAV file that will be played when a study has
been received successfully.
•
Error:
This parameter specifies the Windows.WAV file that will be played when an error has
occurred during the receive process for a study.
•
Post Process:
When you need a special process to run, such as for quality control matching or
repairing DICOM tags, enter the Dynamic Link Library here (qc.dll for example), or use the Browse
button to locate the DLL you want to use.
•
Icon Images
•
Play Receive Sound:
This setting controls whether a sound will be played when studies are
received by the system. The sound will be played when an entry is added to the Receive Log.
•
RadWorks 5.1 Receive:
Check this check box to specify whether the RadWorks 5.1 Receive thread is available in the
system. If this check box is disabled, RadWorks 5.1 systems will not be able to send data to this
system in AMI (5.1) format.
-
Directory:
Directory where the Connection Service receives studies from RadWorks 2.x or
higher and RadWorks 1.31 systems that arrive in AMI format.
-
Share:
Share name for the directory (as specified above) that is made available to other
systems that send data to this system. This share is automatically created during the
installation of the software.
-
Pipe:
Pipe name of the interface provided by the Connection Service for other RadWorks 2.x
or higher systems that want to send data to it in AMI format.
-
Minimum Free Space:
This indicates (in megabytes) the minimum amount of free space that
should be kept available when receiving studies. Should the amount of free space come
below the level specified here, the study will be refused.
3-5-5-6
DICOM
The DICOM tab contains a collection of settings that control various DICOM parameters of the
Connection Service.
To view and set these DICOM settings:
1.) Select
Configuration
>
Connection Service...
from the Data Selector menu.
2.) Select the DICOM tab from the Connection Service Properties page.