Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1
Environmental Requirements of System based on Dell Precision 340
Table 2-2
Dell Precision 340 Workstation
Table 2-3
NEC LCD 1880 Monitor
Table 2-4
Global Modem
2-2-1-1
Cooling
The cooling requirement for the LOGIQworks is within the range 1220-2100 BTU/hr. This is depending
on whether a CRT Monitor or LCD Monitor is used, and on the hardware options installed. For detailed
BTU/hr. see table 2-7 and table 2-8. The BTU/hr. value relates to the maximum operational Power
consumption of the LOGIQworks Workstation. This figure does not include cooling needed for lights,
people, or other equipment in the room. Each person in the room places an additional 300 BTU/hr.
Demand on the cooling system.
Operation
Non-Operation
Storage
Transport
Temperature
10 to 35
o
C
(50 to 95
o
F)
-40 to 65
o
C
(- 40 to 149
o
F)
-40 to 65
o
C
(- 40 to 149
o
F)
-40 to 65
o
C
(-40 to 149
o
F)
Humidity
20% to 80%
non-condensing
20% to 80%
non-condensing
20% to 80%
non-condensing
20% to 80%
non-condensing
Altitude
-52 - 10000ft
-16 - 3048 m
-52 - 34777ft
-16 - 10600 m
52 - 34777ft
-16 - 10600 m
52 - 34777ft
-16 - 10600 m
Operation
Non-Operation
Storage
Transport
Temperature
5 to 35
o
C
(41to 95
o
F)
-10 to 60
o
C
(- 14 to 140
o
F)
-10 to 60
o
C
(-14 to 140
o
F)
-10 to 60
o
C
(-14 to 140
o
F)
Humidity
30% to 80%
non-condensing
10% to 85%
non-condensing
10% to 85%
non-condensing
10% to 85
non-condensing
Altitude
0 - 15912ft
0 - 4850m
0 - 45000ft
0 - 13716m
0 - 45000ft
0 - 13716m
0 - 45000ft
0 - 13716m
Operation
Non-Operation
Storage
Transport
Temperature
0 to 50
o
C
(32 to 120
o
F)
0 to 50
o
C
(32 to 120
o
F)
-20 to 70
o
C
(-4 to 158
o
F)
-20 to 70
o
C
(-4 to 158
o
F)
Humidity
20% to 90%
non-condensing
20% to 90%
non-condensing
20% to 90%
non-condensing
20% to 90%
non-condensing
Altitude
N/A
N/A
N/A
N/A