LOGIQ
WORKS
® B
ASIC
S
ERVICE
M
ANUAL
4-4
Section 4-3 - System Check
1.) Enter the Host name or IP address of the system the association shall be tested with.
2.) Enter the DICOM port number of the remote system.
3.) Enter the DICOM Application Entity Title of the remote system.
4.) Then press the PING-button.
Figure 4-1 DICOM Pinger
1.
2.
3.
5.Output
Window
4.