GE M
EDICAL
S
YSTEMS
D
IRECTION
5184024-100, R
EVISION
5
LOGIQ
WORKS
® B
ASIC
S
ERVICE
M
ANUAL
Chapter 3 Installation
3-169
-
Deletion Rules:
Rules which determine whether data will be deleted if it is older than the
specified time frame. These rules are SQL (Structured Query Language) rules. Once one or
more rules are entered, Auto Delete will be activated. If none of the studies match the criteria
that have been set, disk space will not be released and new studies will be rejected due to a
lack of disk space. Exercise caution when setting these rules.
Deletion rules are only executed during a maintenance run.
RadWorks 6.1 Services must be up and running at the time set for a maintenance run.
Deletion rules are case sensitive.
•
Disk space deletion
-
Type in the
Local Drives
field to define the drive letter and the minimum and maximum
amount of free disk space (in percentages). Disk space values represent the percentage of
free disk space available. For example, the setting 'C: 50 70' means that if the C: drive is more
than 70% full, files will be deleted (according to the set of rules) until the 50% or less mark is
reached. If the free space is less than the maximum percentage specified, the Database
Service will commence deleting data until the minimum size is reached (following the deletion
rules).
-
Type in the
Deletion Rules
field to add rules which determine if data will be deleted if disk
space is above the maximum value. These are SQL rules. Automatic deletion will be activated
once one or more rules have been entered. If Time to Wait or Date Stamp to Use have also
been set, these rules will be executed first.
If none of the studies match the criteria that have been set, disk space will not be released
and new studies will be rejected due to lack of disk space. Exercise caution when setting
these rules.
Deletion rules are only executed during a maintenance run.
RadWorks Services must be up and running at the time set for a maintenance run.
Deletion rules are case sensitive.
3-5-7-4
Input
To specify the input-related settings for the Database Service:
1.) Select
Configuration
>
Database Service...
from the Data Selector menu.
2.) Select the Input tab from the Database Service Properties page.
Содержание H49021AA
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION 5184024 100 REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL vii ...
Страница 418: ...GE MEDICAL SYSTEMS DIRECTION 5184024 100 REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL II ...
Страница 419: ......