Chapter 3 Installation
3-183
3-5-10
Connectivity (Application Level)
3-5-10-1
Setting up and configuring origins
When your system is connected to a network, any properly equipped DICOM compliant system
connected to that network could query your system’s database and even send studies. Access to your
system can be restricted using “origins”. When another system tries to communicate with your
LOGIQworks, you can tell your system to only allow communication if this origin is one of a pre-
configured set of permitted origins.
1.) From the Data Selector menu, select
Connection -> Origins...
2.) Press the Add Button, the Origin Properties page is displayed.
NOTE
For Multivender Philips HD11 and Philips HD11 XE, 3D works for multiframes only if the field "Time
Frame Vector" is disabled on the Philips scanner. For details refer to the Philips manual.
Figure 3-156 Configure Origins (Default)