Chapter 3 Installation
3-139
•
Startup Delay (sec.):
Specifies the delay after start-up when the first maintenance check will be
performed. (Default: 30)
•
Maintenance Times:
Specifies at which moments in time a maintenance check will be performed.
These checks are done every day, and only if the system is up and running at that. (Default: 01:00,
09:00 and 17:00).
*
You can add a time by filling in the field beneath the list of Maintenance Times and clicking
on the
Add
button.
*
You can delete a time by selecting the time in the Maintenance Times list and clicking the
Remove
button.
*
You replace one maintenance time with another by filling in the field beneath the list of
Maintenance Times, selecting the time you want to replace and clicking the
Replace
button.
•
Max. No. Send Log Records:
Maximum number of Send Log records. This number is used when
maintenance becomes active. If the number of records in the Send Log exceeds the specified
number, the oldest records will be deleted from the Send Log until the required maximum is
reached. Setting this number to zero (as shown) will disable maintenance for the Send Log.
(Default: 500)
Figure 3-123 Connection Service Properties - Maintenance (Default Settings)
Administrator only
Administrator only