GE M
EDICAL
S
YSTEMS
D
IRECTION
5184024-100, R
EVISION
5
LOGIQ
WORKS
B
ASIC
S
ERVICE
M
ANUAL
3-58
Section 3-4 - Completing the Installation
3-4-3-18
Global Modem Connections
1.) To do Modem RS232 Connection, plug one end of the RS232 cable into the RS232 connector on
the modem, and the other end into a serial port connector on your computer, Port 1.
2.) To do Modem Line Connection, plug one end of the telephone wire into the modem’s LINE jack,
and the other end into a phone line wall jack.
3.) To do Modem Power Connection:
a.) Plug the female end of the Power Cable into the AC/DC Convertor and the male end into a
vacant AC power outlet on the wall.
b.) Plug the AC/DC convertor cable into the PWR jack on the modem.
Figure 3-45 Global Modem Signal and Power connections for Dell Precision 340
NOTE
The modem delivery includes a large set of global telephone standards adapters for the telephone wire.
NOTE
The LINE jack is not interchangeable with the PHONE jack. Do not plug the phone into the LINE jack
or the line cable into the PHONE jack.
CAUTION
Use only the
AC
/DC power supply supplied with the modem. Use of any other
AC
/DC power supply will
void the warranty and could damage the modem.
Telephone Wire
Modem
Phone line wall jack
Power
Cable
AC/DC Convertor
RS232
Cable
076D2610
Power
Cable,
female
connector
Power Cable,
male
connector
To PC Com 13