3 - 8
Operation
K0260 Issue No. 8
Control Mode
When the system is in measure mode, to enter control mode press:
The display changes to:
Note:
The number of parameters displayed depends on the settings made in set-up, see previous page.
New set-point
To change the Aim value, press the required parameter key and, using the numeric keys, set the
new Aim value. If necessary, use
DELETE
to remove the last digit set in the Aim value display field.
The display shows each numeric key press, a beep sounds with each key press. When the display
shows the new Aim value press
ENTER
.
The display shows the parameter changing as the system controls to the set-point, at the set rate
of change.