Druck ADTS 505 User Manual
6 - 7
K0260 Issue No. 8
6
HELP
•
The HELP key provides information on each key function. The help message generally gives
associated functions and ways of changing the use of each key. When the display shows the
main pressure display, the user can get help on any of the keys on the key-pad. To do this:
•
Press the required key for information, then press HELP.
•
Press
CLEAR/QUIT
to exit the help system.
ENTER
•
Press
ENTER
to complete a numeric entry.
CLEAR/QUIT + ENTER (ABORT)
•
When
CLEAR/QUIT
and
ENTER
are pressed simultaneously, the ADTS 505 restarts from the
power-up sequence.
•
When these keys are pressed all output valves immediately close, isolating the aircraft
systems.
•
The
ABORT
function should only be used as a last resort:
•
When a restart is required after the system has shutdown displaying an error message.
•
As an emergency measure.
•
If the system is restarted by pressing
ABORT
, it vents to atmosphere during the power-up
sequence. The aircraft systems remain at the original pressures. At a later stage in the
power-up sequence, the aircraft system pressures are sampled and the Ps (static) and Pt
(pitot) channels reconnected with a small pressure loss.
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...