
Druck ADTS 505 User Manual
6 - 29
K0260 Issue No. 8
6
Altitude Correction
An altitude correction must be made to allow for the
difference in height between the reference level of the
ADTS 505 and the reference level of the aircraft’s
altitude sensors. Select the Instrument Altitude
Correction and, using the numeric keys, enter the
correction value (see Fig 2-2).
Auto leak
Auto leak automatically regains control if the leak rate becomes too high for any of the control channels. Auto
leak operates at preset rates of 3000ft/min and 300knots/min. This facility can be switched off and on in set-
up menu.
Select Test Sequence File
Files stored in the computer
1.
From the main menu, select Custom Test
Sequence File.
2.
Use the Up and Down keys to highlight the test
file.
3.
Use the SELECT key to load the file.
4.
Press MAIN MENU to return to the Main Menu
Screen.
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...