3.7.8 Go To Ground
On completion of testing and, before disconnecting from the aircraft system or UUT, the pressures
in the system must be taken to the local atmospheric pressure (ground) with zero airspeed.
•
If the display shows "Leak Measure", press
LEAK MEASURE/CONTROL
.
•
Press
GROUND
, (go to ground).
•
The pressure in the system changes towards ground. The ground pressure replaces the
static or altitude aim value; zero replaces the airspeed aim.
Note:
If required, new rates of change can be entered while going to ground.
•
Wait for the display to show zero airspeed and atmospheric pressure.
•
The display changes to "safe at ground" for 5 seconds then to "aircraft safe at ground"
with the message "confirming system settings .....".
Note:
During this operation:
keep the power supply connected and switched on.
the system disables the key-pad.
it is safe to disconnect the pipes from the aircraft system or UUT.
•
After completing this operation the display changes back to "safe at ground" with the
message "Press Clear/Quit to continue".
•
It is now safe to continue testing or to switch off and disconnect the aircraft system or
UUT.
Note:
If the ADTS 505 stays in the "safe at ground" for 30 minutes the
"confirming system settings ....." operation will be repeated. The
operation automatically re-checks the ADTS controller settings,
the pumps start and the display shows:
No pressure is applied to the output ports during this operation.
Shut-down
•
When the airspeed is zero and the Ps channel pressure is close to ground, the
atmospheric pressure is re-measured to update the recorded ground pressure (QFE).
•
Press
QUIT/CLEAR
to continue with normal operation; the ground and zero valves
automatically close.
ALT
Ps
SPEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEASURE
CONTROL
HELP
SETUP
AIRCRAFT
SAFE AT
GROUND
confirming system settings 1 of 4....
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...