Druck ADTS 505 User Manual
6 - 31
K0260 Issue No. 8
6
Creating Custom Test Sequences for the ADTS 505
Instructions for generating or editing source data for test table values
The data shown on each screen or table of the test sequence comes from a single data source file. This .CSV
format file uses numeric data fields separated by commas. Files of this type may be viewed and edited using
a spread sheet program such as Excel. Files would normally be saved with an applicable name for easy
identification and selection.
A template file (TEMPLATE.CSV), provided on the software distribution with this product, may be used as the
basis for creating new aircraft test sequences. An example, below, shows how the data fields of the file can
be used in the test screens.
F
IGURE
6-11 C
USTOMER
T
EST
S
EQUENCE
F
ILE
S
ELECTIONS
Test Table Title
Rate Units
Aim Units
Header 2
Header 1
Ref
Test Descriptor
Message
Ps, Pt Aims
Ps, Pt Rates
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...